Page 26

744

HART Mode Users Guide

2.Use the uand dkeys to select the Sensor Trim operation, and press e.

3.Follow the instructions on the display. For pressure, there is a message to connect a pressure module to the calibrator, and another to press the ckey to zero the pressure module.

4.Press Continue.

pe18s.bmp

5.Enter the trim value you want the calibrator to produce, and press e, or in the case of a pressure transmitter (as above), press Fetch to load the value being measured by the Fluke pressure module into the dialog box.

6.Press Trim.

7.If prompted to do so, repeat steps 2 and 3, following the prompts on the display, to trim the remaining points.

Note

Select trim points near the range limits unless the transmitter’s documentation specifies otherwise.

If you get an error message that for the selected trim point there is an excess correction, sometimes you can solve the problem by doing a series of Sensor Trims in small steps from the previous trim point toward the desired trim point.

18

Image 26
Contents 744 Page Table of Contents Hart Mode Users Guide 744List of Tables 744 List of Figures 744 Documenting Process Calibrator Hart Mode IntroductionHow to Contact Fluke Connecting to a Hart Transmitter ∙ Poll address if not Documenting Process Calibrator Hart Mode Supported vs. Generic Transmitters Supported vs. Generic Transmitters Viewing Process Variables Setup OperationsCommunication Operations Hart Mode Menu Tree Sensor Hart Output Hart Information Abort Softkey Interaction between Analog Mode and Hart ModeService Operations Hart Mode Menus for Adjustment Understanding Hart CalibrationBlock Diagram of a Hart Transmitter Calibrating a Supported Hart Transmitter Loop Test Output Trim Sensor TrimPress Continue Cloning a Transmitter 744 Index Device identification