Fluke 744 manual Cloning a Transmitter

Page 27

Documenting Process Calibrator HART Mode

Cloning a Transmitter

Cloning a Transmitter

Cloning copies the Basic Setup information from one transmitter to another. You can clone generic and supported transmitters.

To use the calibrator to clone a transmitter:

1.Connect the calibrator to the transmitter you want to clone. You only need a communication connection.

2.From the Active Device screen, press the Setup softkey.

3.Use the uor dkeys to select Basic from the list on the screen and press e.

4.Press the Store Page softkey to load the settings into the calibrator’s memory (not the transmitter).

5.Disconnect the calibrator from the transmitter and connect it to the transmitter you want to configure. Press the Abort softkey to establish communication.

6.Press the Setup softkey, then select Basic to return to the Basic Setup screen.

7.Press the Recall Page softkey to refresh the calibrator display with the cloned parameter settings. The settings are not transmitted to the transmitter at this time.

8.Use the uor dand ekeys to select parameters one at a time, change them, or leave them as cloned, and press the Send softkey to transmit them to the transmitter.

pe09s.bmp

19

Image 27
Contents 744 Page Table of Contents 744 Hart Mode Users GuideList of Tables 744 List of Figures 744 Introduction Documenting Process Calibrator Hart ModeHow to Contact Fluke Connecting to a Hart Transmitter ∙ Poll address if not Documenting Process Calibrator Hart Mode Supported vs. Generic Transmitters Supported vs. Generic Transmitters Setup Operations Communication OperationsViewing Process Variables Hart Mode Menu Tree Sensor Hart Output Hart Information Interaction between Analog Mode and Hart Mode Service OperationsAbort Softkey Understanding Hart Calibration Hart Mode Menus for AdjustmentBlock Diagram of a Hart Transmitter Calibrating a Supported Hart Transmitter Loop Test Sensor Trim Output TrimPress Continue Cloning a Transmitter 744 Index Device identification