Fluke 744 manual Output Trim, Sensor Trim

Page 25

Documenting Process Calibrator HART Mode

Calibrating a Supported HART Transmitter

Output Trim

An Output Trim adjusts the tranmitter’s Output stage. You can perform an Output Trim on both generic and supported transmitters.

To perform an Output Trim:

1.From the Active Device screen Press the Service softkey, followed by Output Trim.

Sensor Trim

You can only perform a Sensor Trim on a supported transmitter. A Sensor Trim adjusts the Input stage of a HART transmitter. If your application does not use the Output stage of the transmitter, Sensor Trim is the only adjustment you need to make.

A Sensor Trim may involve one or more trim points. This is controlled by the transmitter’s software. You can only perform a Sensor Trim on a supported transmitter.

To perform a Sensor Trim, proceed as follows:

1.From the Active Device screen, press the Service softkey, followed by Sensor Trim.

 

pe16s.bmp

 

2.

Press Fetch to load the mA value being measured by

 

 

the calibrator into the dialog box.

 

3.

Press Send to trim the low analog output point.

 

4.

To trim the high analog output point, repeat steps 2

pe17s.bmp

 

and 3, following the prompts on the display.

 

17

Image 25
Contents 744 Page Table of Contents 744 Hart Mode Users GuideList of Tables 744 List of Figures 744 Introduction Documenting Process Calibrator Hart ModeHow to Contact Fluke Connecting to a Hart Transmitter ∙ Poll address if not Documenting Process Calibrator Hart Mode Supported vs. Generic Transmitters Supported vs. Generic Transmitters Communication Operations Setup OperationsViewing Process Variables Hart Mode Menu Tree Sensor Hart Output Hart Information Service Operations Interaction between Analog Mode and Hart ModeAbort Softkey Understanding Hart Calibration Hart Mode Menus for AdjustmentBlock Diagram of a Hart Transmitter Calibrating a Supported Hart Transmitter Loop Test Sensor Trim Output TrimPress Continue Cloning a Transmitter 744 Index Device identification