Fluke 744 manual Interaction between Analog Mode and Hart Mode, Service Operations, Abort Softkey

Page 20

744

HART Mode Users Guide

Service Operations

The Service softkey provides access to Loop Test, Pressure Zero Trim (where applicable), Output Trim, and Sensor Trim operations. For generic transmitters, only Loop Test, Output Trim, and Pressure Zero Trim are available. (See Table 1.) The trim (adjustment) operations are described later in this manual.

Note

Loop Test and Output Trim are not available if the transmitter is in multidrop mode.

Interaction between Analog Mode and HART Mode

Analog mode is normal calibrator operation, as described in the 744 Users Manual. HART mode is activated when you press the r(HART) key. This action starts communication over the HART interface.

You can switch between HART and analog modes easily by pressing r(or pressing Mto go to analog mode from HART mode), gaining the benefit of having the transmitter automatically,set up analog mode for appropriate measure and source functions, if desired.

Abort Softkey

The Abort softkey terminates the communication operation underway, and returns control to the previous screen. From the Active Device screen, Abort calls up the browser, in which you can view the list of supported transmitters.

For supported transmitters, the transition to analog mode goes to the MEASURE/SOURCE screen, which provides an easy way to proceed with an “as found” calibration.

For generic transmitters, the transition to analog mode involves choosing the MEASURE or SOURCE screen, from which you select the appropriate function.

For supported or generic transmitters, when you press

rto return to HART mode, the Active Device screen is displayed. The HART serial communication connection remains active as you switch between HART and analog modes.

12

Image 20
Contents 744 Page Table of Contents Hart Mode Users Guide 744List of Tables 744 List of Figures 744 Documenting Process Calibrator Hart Mode IntroductionHow to Contact Fluke Connecting to a Hart Transmitter ∙ Poll address if not Documenting Process Calibrator Hart Mode Supported vs. Generic Transmitters Supported vs. Generic Transmitters Viewing Process Variables Setup OperationsCommunication Operations Hart Mode Menu Tree Sensor Hart Output Hart Information Abort Softkey Interaction between Analog Mode and Hart ModeService Operations Hart Mode Menus for Adjustment Understanding Hart CalibrationBlock Diagram of a Hart Transmitter Calibrating a Supported Hart Transmitter Loop Test Output Trim Sensor TrimPress Continue Cloning a Transmitter 744 Index Device identification