Toshiba LF414 manual Overview, Features, - 10 -, 6F8A0870, Intelligent functions

Page 11
2. Overview

6F8A0870

2. Overview

The LF414/LF600F, LF414/LF610F, LF414/LF620F and LF414 electromagnetic flowmeter can be use in the following hazardous (classified) locations.

Class, Division 2, Groups A, B, C and D,

Class, Division 2, Groups E, F and G

Class

This product is a converter used for electromagnetic flowmeters that measure the volumetric flow rate of conductive fluid using Faraday's law of electromagnetic induction.

You can bring out the functions of the converter when you place it in the converter housing you prepare and use it in combination with a fluid rate measurement detector.

The converter sends out a signal to drive the detector excitation coil, which generates a magnetic field inside the detector. The converter receives the signal electromotive force obtained by the detector. The signal electromotive force is proportional to the generated flow rate in the fluid using Faraday's law of electromagnetic induction. After carrying out its operation, the converter converts the signal electromotive force to an analog signal output and displays the status, as a flow rate value.

Features

With a linear relationship between the flow rate and output signal, the electromagnetic flowmeter is used as an easy-to-read indicator. In addition to this feature, it has the following outstanding features:

(1)Wide flow velocity range setting, such as a flow velocity range of 0 to 0.3 ft/s and 0 to 32.8 ft/s (0 to 0.1 and 0 to 10m/s) is achieved.

(2)The unique Noise Sentry filter circuit enables you to obtain stable output automatically.

(3)Full graphic LCD that enables display of a large amount of information

With a large amount of a maximum of 14 characters x 8 lines, you can easily set up various displays including bar graphs and alarm indications.

The backlight display allows you to read the indicator easily.

(4) Use of infrared switches

Use of infrared switches allows you to perform various operations, without opening the converter housing cover.

(5)Intelligent functions

The widely used HART protocol communications system is used as a standard feature. This product supports PROFIBUS*2 communication (optionally available).

*1 HART protocol: “HART” stands for Highway Addressable Remote Transducer and is a communication protocol recommended by HCF (HART communication Foundation) for industrial sensors.

*2 PROFIBUS:PROFIBUS, which stands for PROCESS FIELDBUS, is a field bus that is approved by international standard IEC61158. The electromagnetic flowmeter supports PRFIBUS PA for process automation.

10 -

Image 11
Contents 6F8A0870 Integral type LF414/LF600F, LF414/LF610F, LF414/LF620F 6F8A0870- 1 - About Safety Precautions- 2 - SAFETY PRECAUTIONSExplanation of signal words Safety symbols„ Do not modify or disassemble the enclosure SAFETY PRECAUTIONS Safety Precautions for Installation and Wiring- 3 - DON’T- 4 - SAFETY PRECAUTIONS continuedSafety Precautions for Installation and Wiring DON’T- 5 - Safety Precautions for Maintenance and InspectionWarranty and Limitation of Liability Usage limitation2 Wire cables correctly and securely Handling Precautions- 6 - ・Places where excessive vibration or mechanical shock occurs- 7 - Handling Precautions continued- 8 - Table of Contents- 9 - 1. Product Inspection and Storage1.1 Product Inspection 1.2 Strage5 Intelligent functions Features2. Overview - 10 -Integral 1 Meter size of 1/2 inch 15mm 3. Names of Parts3.1 Appearance 3.1.1 Appearance of LF414/LF600F, LF414/LF610FGround terminal for converter 2 Meter size of 1inch 25mm- 12 - Figure 3.1.2 Appearance of LF414/LF600F, LF610F Meter size 1 inch25mmMeter size 1 1/2 to 4 inch 40 to 100mm 3 Meter size of 1 1/2 to 4 inch 40 to 100mm- 13 - Figure 3.1.3 Appearance of LF414/LF600F, LF610Fprovided 4 Meter size of 6 and 8 inch 150 and 200mm- 14 - Lifting LugsGround terminal for converter Flow direction arrow Detector 3.1.2 Appearance of LF414/LF620F- 15 - Figure 3.1.5 Appearance of LF414/LF620F Meter size 1/2inch15mmGround terminal for converter Flow direction arrow - 16 -Figure 3.1.6 Appearance of LF414/LF620F Meter size 1 inch25mm Figure 3.1.7 Appearance of LF414/LF620F - 17 -Ground terminal for converter Lifting Lugs - 18 -Figure 3.1.8 Appearance of LF414/LF620F Meter size 6 and 8 inch 150 and 200mmSignal cable 3.1.3 Appearance of LF414Separate 1 Meter size of 1/2 inch 15mm - 19 -for Excitation cable - 20 -for Signal cable Figure 3.1.10 Appearance of LF414 Meter size 1 inch25mmExcitation cable - 21 -Signal cable Figure 3.1.11 Appearance of LF414Meter size 6 and 8 inch150 and 200mm - 22 -Figure 3.1.12 Appearance of LF414 - 23 - Signal cable terminal3.2 Construction of the terminal blocks 3.2.2 Terminal Block Construction of LF414 Type„ Use an appropriate device to carry and install 4. InstallationSafety Precautions for Installation - 24 -2. Avoid places where excessive pipe vibration occurs 4.1 Notes on Selecting the Installation Location4.2 Mounting Procedure 4.2.1 Pipe checks - 25 -Figure 4.3 Model Diagram of Unsupported Pipes - 26 -Figure 4.2 Example of Pipe Fixing Procedure Figure 4.4 LF414 flowmeter detector piping connections 4.2.2 Installation Procedure- 27 - Table 4.1 Bolt length and Nut tightening torque - 28 -L=5D 1a Ideal Upstream Straight Pipe Length Installation Requirements1b Optional “Mount Anywhere” Installation Mount-Anywhere Technology 4.3 Piping ConnectionsFigure 4.5 Detector Piping Orientation 2 Pipe Orientation- 30 - Figure 4.6 Installation position of the detectorFigure 4.7 Flow direction arrow on the detector 3 Flow Direction4 Preventing an Empty Pipe Condition - 31 -Figure 4.10 Grounding the LF414/LF620F Type 4.4 Grounding1 Grounding of the LF414/LF620F type - 32 -- 33 - 2 Grounding of the LF414 type- 34 - 5.WiringSafety Precautions for Wiring independently from power equipment. 100 ohm 6F8A0870- 35 - „ Install a switch and fuse to isolate theTable 5.1 Installation Cables Notes on wiring5.1 Cables - 36 -- 37 - 5.3 Notes on Wiring 5.3.1 Notes on Instrumentation-Converter Wiring5.3.2 Notes on Wiring of the Separate type LF414 5.2 External Device Connections and Grounding- 38 - 5.4 Wiring 5.4.1 Terminal Treatment of Cables1 Power cable, current output cable, digital I/O cables Separate 2 Excitation cableFigure 5.3 Terminal Treatment of Flow Rate Signal Cable 3 Connecting the input signal cableNotes on signal cable shield processing work - 39 -Figure 5.5 Connecting a Cable to Terminal Block 5.4.2 Cable Connection- 40 - „ Check the wiring between the converter and related instruments 6. OperationPreparatory check - 41 -„ Do not conduct wiring work when power is applied 7. Maintenance and TroubleshootingSafety precaution for Maintenance and Troubleshooting - 42 -materials 7.1 Maintenance„ Cleaning Check/Replacement of the display unitSTART Are power supplies correct NO for each device? YES 7.2 Troubleshooting7.2.1 Flow rate is not indicated - 44 -Clean the inside of the detector pipe 7.2.2 Flow rate indicated is not correct- 45 - Perform the zero adjustment Refer to combined converters manualUse a power supply within the specified range 7.2.3 Flow rate indication is not stable- 46 - Is power supply voltageSquare-Wave Excitation Figure 8.1 Principle of Operation 8. Principle of Operation- 47 - Measurement range in terms of flow velocity 9. Specifications9.1Specifications - 48 -Figure 9.1 Electrical Conductivity vs. Cable Length - 49 -Electrical conductivity μS/cm Table 9.1 Standard Flow Range - 50 -Unit m3/h - 51 -Table 9.2 Flow velocity vs. flow volume Unit gal/min9.2 Type Specification Code Table 9.3 Type Specification Code- 52 - Exciting cable length Nominal cross-sectional area Overall diameter Table 9.4 Type Specification Code Exciting Cable and Signal Cable- 53 - 2 Meter size of 1 inch 25mm 10. Outline Dimensions10.1 Outline dimensions of LF414/LF600F, LF414/LF610F 1 Meter size of 1/2 inch 15mm- 55 - 3 Meter size of 1 1/2 to 8 inch 40 to 200mmWaight approx 9 lb 4kg 10.2 Outline dimensions of LF414/LF620F- 56 - - 57 - Waight approx 7 lb 3kg 10.3 Outline dimensions of LF414- 58 - - 59 - - 60 - Appendix 1-1 A system block diagram for LF414/LF600F, LF414/LF610F- 61 - 1-2 A system block diagram for LF414/LF620F- 62 - 2-1 A system block diagram for LF414Product code - 63 -Distributor Address Name Phone number (FCF50017