Contents6F8A0870- 1 -
6F8A0870About Safety Precautions
Integral type LF414/LF600F, LF414/LF610F, LF414/LF620FExplanation of signal words
SAFETY PRECAUTIONSSafety symbols
- 2 -- 3 -
SAFETY PRECAUTIONS Safety Precautions for Installation and WiringDON’T
Do not modify or disassemble the enclosureSafety Precautions for Installation and Wiring
SAFETY PRECAUTIONS continuedDON’T
- 4 -Warranty and Limitation of Liability
Safety Precautions for Maintenance and InspectionUsage limitation
- 5 -- 6 -
Handling Precautions・Places where excessive vibration or mechanical shock occurs
2 Wire cables correctly and securely- 7 -
Handling Precautions continued- 8 -
Table of Contents1.1 Product Inspection
1. Product Inspection and Storage1.2 Strage
- 9 -2. Overview
Features- 10 -
5 Intelligent functions3.1 Appearance
3. Names of Parts3.1.1 Appearance of LF414/LF600F, LF414/LF610F
Integral 1 Meter size of 1/2 inch 15mm- 12 -
2 Meter size of 1inch 25mmFigure 3.1.2 Appearance of LF414/LF600F, LF610F Meter size 1 inch25mm
Ground terminal for converter- 13 -
3 Meter size of 1 1/2 to 4 inch 40 to 100mmFigure 3.1.3 Appearance of LF414/LF600F, LF610F
Meter size 1 1/2 to 4 inch 40 to 100mm- 14 -
4 Meter size of 6 and 8 inch 150 and 200mmLifting Lugs
provided- 15 -
3.1.2 Appearance of LF414/LF620FFigure 3.1.5 Appearance of LF414/LF620F Meter size 1/2inch15mm
Ground terminal for converter Flow direction arrow DetectorFigure 3.1.6 Appearance of LF414/LF620F Meter size 1 inch25mm
- 16 -Ground terminal for converter Flow direction arrowFigure 3.1.7 Appearance of LF414/LF620F
- 17 -Figure 3.1.8 Appearance of LF414/LF620F
- 18 -Meter size 6 and 8 inch 150 and 200mm
Ground terminal for converter Lifting LugsSeparate 1 Meter size of 1/2 inch 15mm
3.1.3 Appearance of LF414- 19 -
Signal cablefor Signal cable
- 20 -Figure 3.1.10 Appearance of LF414 Meter size 1 inch25mm
for Excitation cableSignal cable
- 21 -Figure 3.1.11 Appearance of LF414
Excitation cableFigure 3.1.12 Appearance of LF414
- 22 -Meter size 6 and 8 inch150 and 200mm3.2 Construction of the terminal blocks
Signal cable terminal3.2.2 Terminal Block Construction of LF414 Type
- 23 -Safety Precautions for Installation
4. Installation- 24 -
Use an appropriate device to carry and install4.2 Mounting Procedure 4.2.1 Pipe checks
4.1 Notes on Selecting the Installation Location- 25 -
2. Avoid places where excessive pipe vibration occursFigure 4.2 Example of Pipe Fixing Procedure
- 26 -Figure 4.3 Model Diagram of Unsupported Pipes- 27 -
4.2.2 Installation ProcedureFigure 4.4 LF414 flowmeter detector piping connectionsTable 4.1 Bolt length and Nut tightening torque
- 28 -1b Optional “Mount Anywhere” Installation Mount-Anywhere Technology
1a Ideal Upstream Straight Pipe Length Installation Requirements4.3 Piping Connections
L=5D- 30 -
2 Pipe OrientationFigure 4.6 Installation position of the detector
Figure 4.5 Detector Piping Orientation4 Preventing an Empty Pipe Condition
3 Flow Direction- 31 -
Figure 4.7 Flow direction arrow on the detector1 Grounding of the LF414/LF620F type
4.4 Grounding- 32 -
Figure 4.10 Grounding the LF414/LF620F Type- 33 -
2 Grounding of the LF414 typeSafety Precautions for Wiring
5.Wiring- 34 -- 35 -
6F8A0870 Install a switch and fuse to isolate the
independently from power equipment. 100 ohm5.1 Cables
Notes on wiring- 36 -
Table 5.1 Installation Cables5.3.2 Notes on Wiring of the Separate type LF414
5.3 Notes on Wiring 5.3.1 Notes on Instrumentation-Converter Wiring5.2 External Device Connections and Grounding
- 37 -1 Power cable, current output cable, digital I/O cables
5.4 Wiring 5.4.1 Terminal Treatment of CablesSeparate 2 Excitation cable
- 38 -Notes on signal cable shield processing work
3 Connecting the input signal cable- 39 -
Figure 5.3 Terminal Treatment of Flow Rate Signal Cable- 40 -
5.4.2 Cable ConnectionFigure 5.5 Connecting a Cable to Terminal BlockPreparatory check
6. Operation- 41 -
Check the wiring between the converter and related instrumentsSafety precaution for Maintenance and Troubleshooting
7. Maintenance and Troubleshooting- 42 -
Do not conduct wiring work when power is applied Cleaning
7.1 MaintenanceCheck/Replacement of the display unit
materials7.2.1 Flow rate is not indicated
7.2 Troubleshooting- 44 -
START Are power supplies correct NO for each device? YES- 45 -
7.2.2 Flow rate indicated is not correctPerform the zero adjustment Refer to combined converters manual
Clean the inside of the detector pipe- 46 -
7.2.3 Flow rate indication is not stableIs power supply voltage
Use a power supply within the specified range- 47 -
8. Principle of OperationSquare-Wave Excitation Figure 8.1 Principle of Operation9.1Specifications
9. Specifications- 48 -
Measurement range in terms of flow velocityElectrical conductivity μS/cm
- 49 -Figure 9.1 Electrical Conductivity vs. Cable LengthTable 9.1 Standard Flow Range
- 50 -Table 9.2 Flow velocity vs. flow volume
- 51 -Unit gal/min
Unit m3/h- 52 -
Table 9.3 Type Specification Code9.2 Type Specification Code- 53 -
Table 9.4 Type Specification Code Exciting Cable and Signal CableExciting cable length Nominal cross-sectional area Overall diameter10.1 Outline dimensions of LF414/LF600F, LF414/LF610F
10. Outline Dimensions1 Meter size of 1/2 inch 15mm
2 Meter size of 1 inch 25mm- 55 -
3 Meter size of 1 1/2 to 8 inch 40 to 200mm- 56 -
10.2 Outline dimensions of LF414/LF620FWaight approx 9 lb 4kg- 57 -- 58 -
10.3 Outline dimensions of LF414Waight approx 7 lb 3kg- 59 -- 60 -
Appendix 1-1 A system block diagram for LF414/LF600F, LF414/LF610F- 61 -
1-2 A system block diagram for LF414/LF620F- 62 -
2-1 A system block diagram for LF414Distributor Address Name
- 63 -Phone number (
Product codeFCF50017