Toshiba LF414 manual - 50 -, 6F8A0870, 1 Standard Flow Range

Page 51
Table 9.1 Standard Flow Range

6F8A0870

Flow and calibration velocity range:

It calibration by standard Range shown in the table below when Range is not specified.

It calibration when there is specification by flowing quantity Range in which the customer is specified. Is this specification Range flowing quantity of Table 9.1. Please confirm becoming in the upper bound value from the flow velocity chart.

Table 9.1 Standard Flow Range

Meter size

 

Standard flow range

Flow rate

 

Flow rate

Flow velocity

Inch (mm)

 

(gal/min)

 

(m³/h)

(m/s)

 

 

1/2 (15)

25

 

2

3.144

1 (25)

75

 

6

3.395

1 1/2 (40)

175

 

15

3.316

2 (50)

300

 

25

3.537

3 (80)

650

 

60

3.316

4 (100)

1,000

 

100

3.537

6 (150)

2,500

 

200

3.144

8 (200)

4,500

 

300

2.653

Note: The unit of "gal/min" is not exchanged (converted) by “m3/h”

- 50 -

Image 51
Contents 6F8A0870 Integral type LF414/LF600F, LF414/LF610F, LF414/LF620F 6F8A0870- 1 - About Safety Precautions- 2 - SAFETY PRECAUTIONSExplanation of signal words Safety symbols„ Do not modify or disassemble the enclosure SAFETY PRECAUTIONS Safety Precautions for Installation and Wiring- 3 - DON’T- 4 - SAFETY PRECAUTIONS continuedSafety Precautions for Installation and Wiring DON’T- 5 - Safety Precautions for Maintenance and InspectionWarranty and Limitation of Liability Usage limitation2 Wire cables correctly and securely Handling Precautions- 6 - ・Places where excessive vibration or mechanical shock occurs- 7 - Handling Precautions continued- 8 - Table of Contents- 9 - 1. Product Inspection and Storage1.1 Product Inspection 1.2 Strage5 Intelligent functions Features2. Overview - 10 -Integral 1 Meter size of 1/2 inch 15mm 3. Names of Parts3.1 Appearance 3.1.1 Appearance of LF414/LF600F, LF414/LF610FGround terminal for converter 2 Meter size of 1inch 25mm- 12 - Figure 3.1.2 Appearance of LF414/LF600F, LF610F Meter size 1 inch25mmMeter size 1 1/2 to 4 inch 40 to 100mm 3 Meter size of 1 1/2 to 4 inch 40 to 100mm- 13 - Figure 3.1.3 Appearance of LF414/LF600F, LF610Fprovided 4 Meter size of 6 and 8 inch 150 and 200mm- 14 - Lifting LugsGround terminal for converter Flow direction arrow Detector 3.1.2 Appearance of LF414/LF620F- 15 - Figure 3.1.5 Appearance of LF414/LF620F Meter size 1/2inch15mm- 16 - Figure 3.1.6 Appearance of LF414/LF620F Meter size 1 inch25mmGround terminal for converter Flow direction arrow Figure 3.1.7 Appearance of LF414/LF620F - 17 -Ground terminal for converter Lifting Lugs - 18 -Figure 3.1.8 Appearance of LF414/LF620F Meter size 6 and 8 inch 150 and 200mmSignal cable 3.1.3 Appearance of LF414Separate 1 Meter size of 1/2 inch 15mm - 19 -for Excitation cable - 20 -for Signal cable Figure 3.1.10 Appearance of LF414 Meter size 1 inch25mmExcitation cable - 21 -Signal cable Figure 3.1.11 Appearance of LF414- 22 - Figure 3.1.12 Appearance of LF414Meter size 6 and 8 inch150 and 200mm - 23 - Signal cable terminal3.2 Construction of the terminal blocks 3.2.2 Terminal Block Construction of LF414 Type„ Use an appropriate device to carry and install 4. InstallationSafety Precautions for Installation - 24 -2. Avoid places where excessive pipe vibration occurs 4.1 Notes on Selecting the Installation Location4.2 Mounting Procedure 4.2.1 Pipe checks - 25 -- 26 - Figure 4.2 Example of Pipe Fixing ProcedureFigure 4.3 Model Diagram of Unsupported Pipes 4.2.2 Installation Procedure - 27 -Figure 4.4 LF414 flowmeter detector piping connections Table 4.1 Bolt length and Nut tightening torque - 28 -L=5D 1a Ideal Upstream Straight Pipe Length Installation Requirements1b Optional “Mount Anywhere” Installation Mount-Anywhere Technology 4.3 Piping ConnectionsFigure 4.5 Detector Piping Orientation 2 Pipe Orientation- 30 - Figure 4.6 Installation position of the detectorFigure 4.7 Flow direction arrow on the detector 3 Flow Direction4 Preventing an Empty Pipe Condition - 31 -Figure 4.10 Grounding the LF414/LF620F Type 4.4 Grounding1 Grounding of the LF414/LF620F type - 32 -- 33 - 2 Grounding of the LF414 type5.Wiring Safety Precautions for Wiring- 34 - independently from power equipment. 100 ohm 6F8A0870- 35 - „ Install a switch and fuse to isolate theTable 5.1 Installation Cables Notes on wiring5.1 Cables - 36 -- 37 - 5.3 Notes on Wiring 5.3.1 Notes on Instrumentation-Converter Wiring5.3.2 Notes on Wiring of the Separate type LF414 5.2 External Device Connections and Grounding- 38 - 5.4 Wiring 5.4.1 Terminal Treatment of Cables1 Power cable, current output cable, digital I/O cables Separate 2 Excitation cableFigure 5.3 Terminal Treatment of Flow Rate Signal Cable 3 Connecting the input signal cableNotes on signal cable shield processing work - 39 -5.4.2 Cable Connection - 40 -Figure 5.5 Connecting a Cable to Terminal Block „ Check the wiring between the converter and related instruments 6. OperationPreparatory check - 41 -„ Do not conduct wiring work when power is applied 7. Maintenance and TroubleshootingSafety precaution for Maintenance and Troubleshooting - 42 -materials 7.1 Maintenance„ Cleaning Check/Replacement of the display unitSTART Are power supplies correct NO for each device? YES 7.2 Troubleshooting7.2.1 Flow rate is not indicated - 44 -Clean the inside of the detector pipe 7.2.2 Flow rate indicated is not correct- 45 - Perform the zero adjustment Refer to combined converters manualUse a power supply within the specified range 7.2.3 Flow rate indication is not stable- 46 - Is power supply voltage8. Principle of Operation - 47 -Square-Wave Excitation Figure 8.1 Principle of Operation Measurement range in terms of flow velocity 9. Specifications9.1Specifications - 48 -- 49 - Electrical conductivity μS/cmFigure 9.1 Electrical Conductivity vs. Cable Length Table 9.1 Standard Flow Range - 50 -Unit m3/h - 51 -Table 9.2 Flow velocity vs. flow volume Unit gal/minTable 9.3 Type Specification Code - 52 -9.2 Type Specification Code Table 9.4 Type Specification Code Exciting Cable and Signal Cable - 53 -Exciting cable length Nominal cross-sectional area Overall diameter 2 Meter size of 1 inch 25mm 10. Outline Dimensions10.1 Outline dimensions of LF414/LF600F, LF414/LF610F 1 Meter size of 1/2 inch 15mm- 55 - 3 Meter size of 1 1/2 to 8 inch 40 to 200mm10.2 Outline dimensions of LF414/LF620F - 56 -Waight approx 9 lb 4kg - 57 - 10.3 Outline dimensions of LF414 - 58 -Waight approx 7 lb 3kg - 59 - - 60 - Appendix 1-1 A system block diagram for LF414/LF600F, LF414/LF610F- 61 - 1-2 A system block diagram for LF414/LF620F- 62 - 2-1 A system block diagram for LF414Product code - 63 -Distributor Address Name Phone number (FCF50017