Notes on wiring

6F8A0870

Notes on wiring

5.1 Cables CAUTION

(1)Select the cable runs away from electrical equipment (motors, transformers, or radio transmitters) which causes electromagnetic or electrostatic interference.

(2)Deterioration of flowmeter circuit insulation occurs if the converter interior or cable ends get wet or humidified. This in turn causes malfunction of flowmeter or noise problems. Avoid a rainy day if the flowmeter is to be installed outdoors. Even indoors, prevent water from splashing over the flowmeter. Try to finish the wiring as quickly as possible

(3)The converter has an arrestor installed inside. Therefore, do not conduct a withstand voltage test for the converter. To check the insulation of the converter, use a voltage of 250Vdc or less.

(4)After wiring, be sure to install the terminal block protection cover.

(5)Because the excitation cable and flow rate signal cable transmit very delicate signals, pass each of

Separate

them separately through a thick steel conduit tube, keep them away from the large current wiring as

far as possible, and do not install them in parallel.

 

5.1 Cables

Separate

Separate

Use the kind of cables shown in Table 5.1 to wire the converter.

Table 5.1 Installation Cables

 

 

Nominal

Finished

 

Name

Cable name

cross-section

outer

Description

 

 

al area

diameter

 

 

 

 

 

 

Power cable

3-core vinyl sheathed cable or 2-core vinyl

2 mm²

11~13mm

CVV JIS C 3401 or

sheathed cable

equivalent

Output

The number of conductors the cable contains differs depending on the

CVV-S JIS -258-C or

specification of the output signal cable.

 

 

signal cable

Use a shielded cable of finished outer diameter 11 to

13mm and

equivalent

 

nominal cross-sectional area 1.25mm2.

 

 

 

Flow rate

2-core shielded chloroprene cabtyre cable

0.75 mm2

 

2PNCT-S

11~13mm

JIS C 3327 or

signal cable

( Rubber covered cable )

 

 

equivalent

 

 

 

 

 

 

 

 

Excitation

3-core chloroprene cabtyre cable

2 mm2

 

2PNCT

cable

( Rubber covered cable )

1.25 m2

11~13mm

JIS C 3327 or

 

 

 

 

equivalent

- 36 -

Image 37
Contents 6F8A0870 - 1 - 6F8A0870About Safety Precautions Integral type LF414/LF600F, LF414/LF610F, LF414/LF620FExplanation of signal words SAFETY PRECAUTIONSSafety symbols - 2 -- 3 - SAFETY PRECAUTIONS Safety Precautions for Installation and WiringDON’T „ Do not modify or disassemble the enclosureSafety Precautions for Installation and Wiring SAFETY PRECAUTIONS continuedDON’T - 4 -Warranty and Limitation of Liability Safety Precautions for Maintenance and InspectionUsage limitation - 5 -- 6 - Handling Precautions・Places where excessive vibration or mechanical shock occurs 2 Wire cables correctly and securely- 7 - Handling Precautions continued- 8 - Table of Contents1.1 Product Inspection 1. Product Inspection and Storage1.2 Strage - 9 -2. Overview Features- 10 - 5 Intelligent functions3.1 Appearance 3. Names of Parts3.1.1 Appearance of LF414/LF600F, LF414/LF610F Integral 1 Meter size of 1/2 inch 15mm- 12 - 2 Meter size of 1inch 25mmFigure 3.1.2 Appearance of LF414/LF600F, LF610F Meter size 1 inch25mm Ground terminal for converter- 13 - 3 Meter size of 1 1/2 to 4 inch 40 to 100mmFigure 3.1.3 Appearance of LF414/LF600F, LF610F Meter size 1 1/2 to 4 inch 40 to 100mm- 14 - 4 Meter size of 6 and 8 inch 150 and 200mmLifting Lugs provided- 15 - 3.1.2 Appearance of LF414/LF620FFigure 3.1.5 Appearance of LF414/LF620F Meter size 1/2inch15mm Ground terminal for converter Flow direction arrow DetectorFigure 3.1.6 Appearance of LF414/LF620F Meter size 1 inch25mm - 16 -Ground terminal for converter Flow direction arrow Figure 3.1.7 Appearance of LF414/LF620F - 17 -Figure 3.1.8 Appearance of LF414/LF620F - 18 -Meter size 6 and 8 inch 150 and 200mm Ground terminal for converter Lifting LugsSeparate 1 Meter size of 1/2 inch 15mm 3.1.3 Appearance of LF414- 19 - Signal cablefor Signal cable - 20 -Figure 3.1.10 Appearance of LF414 Meter size 1 inch25mm for Excitation cableSignal cable - 21 -Figure 3.1.11 Appearance of LF414 Excitation cableFigure 3.1.12 Appearance of LF414 - 22 -Meter size 6 and 8 inch150 and 200mm 3.2 Construction of the terminal blocks Signal cable terminal3.2.2 Terminal Block Construction of LF414 Type - 23 -Safety Precautions for Installation 4. Installation- 24 - „ Use an appropriate device to carry and install4.2 Mounting Procedure 4.2.1 Pipe checks 4.1 Notes on Selecting the Installation Location- 25 - 2. Avoid places where excessive pipe vibration occursFigure 4.2 Example of Pipe Fixing Procedure - 26 -Figure 4.3 Model Diagram of Unsupported Pipes - 27 - 4.2.2 Installation ProcedureFigure 4.4 LF414 flowmeter detector piping connections Table 4.1 Bolt length and Nut tightening torque - 28 -1b Optional “Mount Anywhere” Installation Mount-Anywhere Technology 1a Ideal Upstream Straight Pipe Length Installation Requirements4.3 Piping Connections L=5D- 30 - 2 Pipe OrientationFigure 4.6 Installation position of the detector Figure 4.5 Detector Piping Orientation4 Preventing an Empty Pipe Condition 3 Flow Direction- 31 - Figure 4.7 Flow direction arrow on the detector1 Grounding of the LF414/LF620F type 4.4 Grounding- 32 - Figure 4.10 Grounding the LF414/LF620F Type- 33 - 2 Grounding of the LF414 typeSafety Precautions for Wiring 5.Wiring- 34 - - 35 - 6F8A0870„ Install a switch and fuse to isolate the independently from power equipment. 100 ohm5.1 Cables Notes on wiring- 36 - Table 5.1 Installation Cables5.3.2 Notes on Wiring of the Separate type LF414 5.3 Notes on Wiring 5.3.1 Notes on Instrumentation-Converter Wiring5.2 External Device Connections and Grounding - 37 -1 Power cable, current output cable, digital I/O cables 5.4 Wiring 5.4.1 Terminal Treatment of CablesSeparate 2 Excitation cable - 38 -Notes on signal cable shield processing work 3 Connecting the input signal cable- 39 - Figure 5.3 Terminal Treatment of Flow Rate Signal Cable- 40 - 5.4.2 Cable ConnectionFigure 5.5 Connecting a Cable to Terminal Block Preparatory check 6. Operation- 41 - „ Check the wiring between the converter and related instrumentsSafety precaution for Maintenance and Troubleshooting 7. Maintenance and Troubleshooting- 42 - „ Do not conduct wiring work when power is applied„ Cleaning 7.1 MaintenanceCheck/Replacement of the display unit materials7.2.1 Flow rate is not indicated 7.2 Troubleshooting- 44 - START Are power supplies correct NO for each device? YES- 45 - 7.2.2 Flow rate indicated is not correctPerform the zero adjustment Refer to combined converters manual Clean the inside of the detector pipe- 46 - 7.2.3 Flow rate indication is not stableIs power supply voltage Use a power supply within the specified range- 47 - 8. Principle of OperationSquare-Wave Excitation Figure 8.1 Principle of Operation 9.1Specifications 9. Specifications- 48 - Measurement range in terms of flow velocityElectrical conductivity μS/cm - 49 -Figure 9.1 Electrical Conductivity vs. Cable Length Table 9.1 Standard Flow Range - 50 -Table 9.2 Flow velocity vs. flow volume - 51 -Unit gal/min Unit m3/h- 52 - Table 9.3 Type Specification Code9.2 Type Specification Code - 53 - Table 9.4 Type Specification Code Exciting Cable and Signal CableExciting cable length Nominal cross-sectional area Overall diameter 10.1 Outline dimensions of LF414/LF600F, LF414/LF610F 10. Outline Dimensions1 Meter size of 1/2 inch 15mm 2 Meter size of 1 inch 25mm- 55 - 3 Meter size of 1 1/2 to 8 inch 40 to 200mm- 56 - 10.2 Outline dimensions of LF414/LF620FWaight approx 9 lb 4kg - 57 - - 58 - 10.3 Outline dimensions of LF414Waight approx 7 lb 3kg - 59 - - 60 - Appendix 1-1 A system block diagram for LF414/LF600F, LF414/LF610F- 61 - 1-2 A system block diagram for LF414/LF620F- 62 - 2-1 A system block diagram for LF414Distributor Address Name - 63 -Phone number ( Product codeFCF50017