Toshiba manual Grounding of the LF414 type, - 33 -, 6F8A0870

Page 34
(2) Grounding of the LF414 type

6F8A0870

(2) Grounding of the LF414 type

Separate

Ground the external grounding terminal of the detector and the FG terminal of the converter (or external grounding terminal of the converter) securely (grounding resistance 100Ω or lower). Use grounding wire material of IV wire 5.5mm2 or more. Do not share a grounding wire with other instruments where grounding current may flow. (An independent grounding is preferable.)

If it is difficult to perform grounding work at the detector side because of a pit installation or other reasons, use a 3-core cable for the excitation cable and connect the E terminal of the detector to the E terminal of the converter. (The E terminal of the converter is internally connected with the FG terminal and the converter case.)

Converter A

G

 

X

 

terminal

B

E

 

Y

FG

block

 

 

 

 

 

 

Shield twisted

 

 

 

 

 

 

 

 

 

 

Grounding terminal

Input signal

 

 

 

 

(100 ohm or less

 

 

 

 

ground resistance)

cable

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Excitation cable

Shield twisted

 

 

 

 

Grounding terminal

 

 

 

 

 

 

 

G

 

E

Converter

 

A

B

X

Y

terminal

 

block

(100 ohm or less

Figure 4.11 (a) ground resistance)

Wiring between Detector and Converter (For grounding the detector, see Figure 4.12 below.)

Converter

A

G

X

 

terminal

B

E

Y

FG

block

Shield twisted

Grounding terminal

(100 ohm or less

ground resistance)

Input signal

 

cable

Excitation cable

 

Shield twisted

Grounding terminal

 

 

 

 

 

 

G

 

E

Converter

 

 

 

 

 

 

terminal

 

A

B

X

 

Y

block

 

 

 

 

 

 

 

Figure 4.11 (b)

Wiring between Detector and Converter (when grounidng of the detector is difficut)

Conductive material piping Example: Metal, etc.

Non-conductive material piping

Example: Resin product or metal piping whose inside is resin lined

Grounding terminal

Grounding

wire

If the piping material is conductive, connect the grounding wires to the both ends of the piping flange.

Grounding terminal

Grounding wire

(100 ohm or less ground resistance)

If the piping material is non-conductive, perform grounding resistance 100Ω or less.

Figure 4.12 Grounding the LF414 Type Detector

- 33 -

Image 34
Contents 6F8A0870 About Safety Precautions 6F8A0870- 1 - Integral type LF414/LF600F, LF414/LF610F, LF414/LF620FSafety symbols SAFETY PRECAUTIONSExplanation of signal words - 2 -DON’T SAFETY PRECAUTIONS Safety Precautions for Installation and Wiring- 3 - „ Do not modify or disassemble the enclosureDON’T SAFETY PRECAUTIONS continuedSafety Precautions for Installation and Wiring - 4 -Usage limitation Safety Precautions for Maintenance and InspectionWarranty and Limitation of Liability - 5 -・Places where excessive vibration or mechanical shock occurs Handling Precautions- 6 - 2 Wire cables correctly and securelyHandling Precautions continued - 7 -Table of Contents - 8 -1.2 Strage 1. Product Inspection and Storage1.1 Product Inspection - 9 -- 10 - Features2. Overview 5 Intelligent functions3.1.1 Appearance of LF414/LF600F, LF414/LF610F 3. Names of Parts3.1 Appearance Integral 1 Meter size of 1/2 inch 15mmFigure 3.1.2 Appearance of LF414/LF600F, LF610F Meter size 1 inch25mm 2 Meter size of 1inch 25mm- 12 - Ground terminal for converterFigure 3.1.3 Appearance of LF414/LF600F, LF610F 3 Meter size of 1 1/2 to 4 inch 40 to 100mm- 13 - Meter size 1 1/2 to 4 inch 40 to 100mmLifting Lugs 4 Meter size of 6 and 8 inch 150 and 200mm- 14 - providedFigure 3.1.5 Appearance of LF414/LF620F Meter size 1/2inch15mm 3.1.2 Appearance of LF414/LF620F- 15 - Ground terminal for converter Flow direction arrow DetectorFigure 3.1.6 Appearance of LF414/LF620F Meter size 1 inch25mm - 16 -Ground terminal for converter Flow direction arrow - 17 - Figure 3.1.7 Appearance of LF414/LF620FMeter size 6 and 8 inch 150 and 200mm - 18 -Figure 3.1.8 Appearance of LF414/LF620F Ground terminal for converter Lifting Lugs- 19 - 3.1.3 Appearance of LF414Separate 1 Meter size of 1/2 inch 15mm Signal cableFigure 3.1.10 Appearance of LF414 Meter size 1 inch25mm - 20 -for Signal cable for Excitation cableFigure 3.1.11 Appearance of LF414 - 21 -Signal cable Excitation cableFigure 3.1.12 Appearance of LF414 - 22 -Meter size 6 and 8 inch150 and 200mm 3.2.2 Terminal Block Construction of LF414 Type Signal cable terminal3.2 Construction of the terminal blocks - 23 -- 24 - 4. InstallationSafety Precautions for Installation „ Use an appropriate device to carry and install- 25 - 4.1 Notes on Selecting the Installation Location4.2 Mounting Procedure 4.2.1 Pipe checks 2. Avoid places where excessive pipe vibration occursFigure 4.2 Example of Pipe Fixing Procedure - 26 -Figure 4.3 Model Diagram of Unsupported Pipes - 27 - 4.2.2 Installation ProcedureFigure 4.4 LF414 flowmeter detector piping connections - 28 - Table 4.1 Bolt length and Nut tightening torque4.3 Piping Connections 1a Ideal Upstream Straight Pipe Length Installation Requirements1b Optional “Mount Anywhere” Installation Mount-Anywhere Technology L=5DFigure 4.6 Installation position of the detector 2 Pipe Orientation- 30 - Figure 4.5 Detector Piping Orientation- 31 - 3 Flow Direction4 Preventing an Empty Pipe Condition Figure 4.7 Flow direction arrow on the detector- 32 - 4.4 Grounding1 Grounding of the LF414/LF620F type Figure 4.10 Grounding the LF414/LF620F Type2 Grounding of the LF414 type - 33 -Safety Precautions for Wiring 5.Wiring- 34 - „ Install a switch and fuse to isolate the 6F8A0870- 35 - independently from power equipment. 100 ohm- 36 - Notes on wiring5.1 Cables Table 5.1 Installation Cables5.2 External Device Connections and Grounding 5.3 Notes on Wiring 5.3.1 Notes on Instrumentation-Converter Wiring5.3.2 Notes on Wiring of the Separate type LF414 - 37 -Separate 2 Excitation cable 5.4 Wiring 5.4.1 Terminal Treatment of Cables1 Power cable, current output cable, digital I/O cables - 38 -- 39 - 3 Connecting the input signal cableNotes on signal cable shield processing work Figure 5.3 Terminal Treatment of Flow Rate Signal Cable- 40 - 5.4.2 Cable ConnectionFigure 5.5 Connecting a Cable to Terminal Block - 41 - 6. OperationPreparatory check „ Check the wiring between the converter and related instruments- 42 - 7. Maintenance and TroubleshootingSafety precaution for Maintenance and Troubleshooting „ Do not conduct wiring work when power is appliedCheck/Replacement of the display unit 7.1 Maintenance„ Cleaning materials- 44 - 7.2 Troubleshooting7.2.1 Flow rate is not indicated START Are power supplies correct NO for each device? YESPerform the zero adjustment Refer to combined converters manual 7.2.2 Flow rate indicated is not correct- 45 - Clean the inside of the detector pipeIs power supply voltage 7.2.3 Flow rate indication is not stable- 46 - Use a power supply within the specified range- 47 - 8. Principle of OperationSquare-Wave Excitation Figure 8.1 Principle of Operation - 48 - 9. Specifications9.1Specifications Measurement range in terms of flow velocityElectrical conductivity μS/cm - 49 -Figure 9.1 Electrical Conductivity vs. Cable Length - 50 - Table 9.1 Standard Flow RangeUnit gal/min - 51 -Table 9.2 Flow velocity vs. flow volume Unit m3/h- 52 - Table 9.3 Type Specification Code9.2 Type Specification Code - 53 - Table 9.4 Type Specification Code Exciting Cable and Signal CableExciting cable length Nominal cross-sectional area Overall diameter 1 Meter size of 1/2 inch 15mm 10. Outline Dimensions10.1 Outline dimensions of LF414/LF600F, LF414/LF610F 2 Meter size of 1 inch 25mm3 Meter size of 1 1/2 to 8 inch 40 to 200mm - 55 -- 56 - 10.2 Outline dimensions of LF414/LF620FWaight approx 9 lb 4kg - 57 - - 58 - 10.3 Outline dimensions of LF414Waight approx 7 lb 3kg - 59 - Appendix 1-1 A system block diagram for LF414/LF600F, LF414/LF610F - 60 -1-2 A system block diagram for LF414/LF620F - 61 -2-1 A system block diagram for LF414 - 62 -Phone number ( - 63 -Distributor Address Name Product codeFCF50017