Orion 2xx manual Specifications, Functional, Performance

Page 28
Magnetostrictive time-of-flight
A return signal is generated from the precise location where the magnetic field of the MLI float intersects the magnetostrictive wire
6 inches to 400 inches (15 to 999 cm)
3-buttonmenu-driven data entry and system security 2-line8-character display
FOUNDATION fieldbus, H1 (31.25 kbits/sec) ITK 4.61
Yes, Device type: Linkmaster 9 VDC
15 mA
0X01
1xRB, 2xAI(s) Adjustable 0-25 seconds

5.3Specifications

5.3.1Functional

System Design

Measurement Principle

Input

Measured Variable

Zero and Span

User Interface

Keypad

Indication

Digital Communication

Interoperability test kit (ITK Revision)

LAS capable

Minimum Operating Voltage

Quiescent Current Draw

DEV Revision

Function Blocks

Damping

Power (Measured at instrument terminals)

Fieldbus General Purpose/Explosion Proof 9 to 32 VDC (17 mA maximum current draw)

FISCO/FNICO

9–17.5 VDC (17 mA maximum current draw)

Housing

 

Material

Aluminum A356T6 (<0.20% copper), optional 316 stainless steel

Cable Entry

34" NPT and M20

5.3.2 Performance

Accuracy

 

 

±0.015"

Repeatability

 

 

±0.005% of full span or 0.005" (0.127 mm) (whichever is greater)

 

 

 

 

Linearity

 

 

0.020% of full span or 0.031" (0.794 mm) (whichever is greater)

Maximum level rate of change

6 inches per second

 

 

 

 

Response time

 

 

0.1 second

Warm-up

 

 

<5 second

 

 

 

 

Upper dead zone

 

 

None

Lower dead zone

 

 

<2" (5 cm), SIL 2: <5" (13 cm)

 

 

 

Ambient temperature range

Transmitter:

-40° to +175° F (-40° to +80° C)

 

 

LCD:

-10° to +160° F (-20° to +70° C)

 

 

 

Process temperature

External Mount:

-40° to +248° F (-40° to +120° C)

 

 

 

-320° to +850° F (-195° to +455° C) (with factory insulated MLI)

 

 

 

 

 

 

Direct Insertion:

-40° to +200° F (-40° to +95° C)

 

High temperature probe:

-40° to +500° F (-40° to +260° C)

 

 

 

Humidity

0 to 99% non-condensing

 

Electromagnetic compliance

 

EN 61326

 

 

Maximum Pressure (Direct Insertion)

1700 psig @ +100° F

 

 

 

(limited to the pressure rating of the selected flange or float)

25

46-649 Jupiter Magnetostrictive Transmitter - FOUNDATION fieldbus™

Image 28
Contents Magnetostrictive Level Transmitter Read this Manual Before Installing Table of Contents Foundation Fieldbus Overview DescriptionDevice Configuration Device Descriptions BenefitsIntrinsic Safety Electrostatic Discharge ESD Handling Procedure InstallationUnpacking Site Preparation Before You BeginEquipment and Tools Configuration Information MountingOperational Considerations ExternalPosition the Jupiter transmitter on the side of the MLI Internal, Direct InsertionWiring Overview Function BlocksUniversal Fieldbus Block Parameters Resource Block Jupiter Magnetostrictive Transmitter Foundation fieldbus Jupiter Magnetostrictive Transmitter Foundation fieldbus Transducer Block Password ParametersTransducer Block Parameters User Calibration Parameters Jupiter Configuration ParametersFactory Parameters Analog Input Block ChannelsFirmware Version AI Block ParametersJupiter Magnetostrictive Transmitter Foundation fieldbus Diagnostic Parameters Simulation Feature Troubleshooting TroubleshootingReference Information Status Messages Display Message Action CommentFF Segment Checklist Size Length TypeAgency specifications Atex Intrinsically safe Agency ApprovalsAgency Drawing Agency Drawing Functional SpecificationsPerformance Physical Inches mmReplacement Parts PartsModel Numbers Enhanced Jupiter Magnetostrictive TransmitterNone, defines external mount Direct Insertion Total Level Float Selection Appendix Transducer Block Parameters ReferencesJupiter Magnetostrictive Transmitter Service Policy