3Com 3644A, 3646A, 3645A manual Store parameters, Save

Models: 3645A 3644A 3646A

1 27
Download 27 pages 29.37 Kb
Page 13
Image 13

3600 SERIES USER MANUAL

it then the output will be changed to OFF status, vice versa.

2.3.4 Store parameters

If you often use a voltage and current as 24V and 2A or 12V and 2.3A and etc, you just need to set up the data at the first time and then store the data in the power supply. When you need it, recall it. 3600 series power supply can store up to 10 sets of parameters.

The parameters include 1) Voltage value; 2) Current value; 3) Maximum voltage; 4) Locked/ unlocked keypad; 5) Maximum power ; 6)Baud rate; 7) Communication address.

The store operation always be done after setting up V-set, I-set etc, the operation is as follows:

Procedure

The operation Methods

LCD display

Step 1

Press “Store” button

ENTER PASSWORD

 

 

 

Step 2

Enter the password ( Or jump to step 4 if the keypad

ENTER PASSWORD

 

is unlocked)

 

 

 

 

 

Step 3

Press “Enter” button ( it will return to step 2 if your

ENTER PASSWORD

 

password is wrong for reentering)

1234

 

 

 

 

 

Step 4

Enter the set value for store number(from 1 to 10) by

SAVE

1

 

using the number key or rotate the rotary knob to

 

 

 

change the set value number

 

 

 

 

 

 

Step 5

Press “Store” button to confirm the set value, if the

SAVE

*

 

number is out of the range from 1 to 10, it will retune to

 

 

 

Step 2 for reenter

 

 

It will exit the store operation at any procedure by press ESC button

For example, set up the voltage=15V, current=2A, Maximum output voltage=18V, keypad locked, Maximum output power=25W, Baud rate=9600, communication address=05, after done the setup , users can store all the above setup as a set of data, such as the 01 or 02 etc set data.

3600 Series Programmable Power Supply

VERSION 2.0 2004.6

11

 

Page 13
Image 13
3Com 3644A, 3646A, 3645A manual Store parameters, Save