Cal in

8-5. Calibration Test Using an External Weight

This function tests the weighing accuracy using an external weight and outputs the result. This is available only when the "GLP output (info)" parameter is set to "1" or "2". (Calibration test does not perform calibration)

When the MC-10K is used

0 .000 g

Press and hold

CC out

Release

Step

1 Connect the AC adapter and warm up the balance

Press

 

CC

0

 

 

 

for at least 30 minutes with nothing on the weighing

1

 

 

 

 

 

 

pan.

 

 

 

 

 

 

 

 

 

Step

2

 

 

key until

 

is

0000 .000 g

 

 

 

Press and hold the CAL

CCout

Select

 

 

 

 

 

 

displayed, then release the key.

 

 

 

 

 

 

Step

3

The balance displays

CC

0 .

 

 

2000 .000 g

 

 

 

 

 

If you want to change the target weight, press

 

 

 

 

 

the

SAMPLE key and go to step 4.

 

 

 

 

 

 

 

 

A list of usable weights is shown on page 19.

 

 

 

 

 

 

 

If you use the target weight value stored in

2000 .000 g

 

 

 

 

 

the balance, go to step 5.

 

 

 

 

 

 

 

Step

4 Specify the target weight value as follows:

 

 

 

 

 

 

SAMPLE key... To switch between the target weight

 

2000 .120 g

 

 

 

 

 

 

selection mode (All of the digits blinking) or

 

 

 

 

 

 

 

 

 

 

 

 

 

 

the value adjustment mode (The selected

 

 

 

 

 

 

 

 

digits blinking).

 

 

 

 

2000 .120 g

 

 

 

 

 

 

With the MC-10K, the digits in the 1st, 2nd,

 

 

 

 

 

 

and 3rd decimal places can be adjusted.

 

 

 

 

 

 

 

 

With the MC-31K, the digits in the 1st and 2nd

2000 .123 g

 

 

 

 

 

 

decimal places can be adjusted.

 

 

 

 

RE-ZEROkey . To select the target weight (All of the digits

 

 

 

CC

0

 

 

 

blinking) or adjust the value (The selected

 

 

 

0 .000

 

 

 

digits blinking). Refer to page 19.

 

Example:

 

 

 

 

 

 

 

 

 

 

PRINT key...... To store the new weight value. Even if the

 

 

 

 

New weight

 

 

 

2 k

 

 

 

AC adapter is removed, the data is

 

2000.123 g

 

 

 

 

 

 

 

maintained in non-volatile memory.

 

 

 

 

 

 

CAL

key ......... To cancel the operation and return to

CC 0 .

 

Place weight

 

Step

5 Confirm that there is nothing on the pan and press the

PRINT

 

 

 

 

 

 

 

key. The balance measures the zero point and displays the

 

 

 

 

 

 

measured value. Do not allow vibration or drafts to affect the

 

 

 

2

 

 

balance. The balance displays the target weight value.

 

 

 

 

Step

6 Place the displayed target weight on the pan and press the

 

 

2000 .123

 

 

PRINT key. The balance measures the target weight and

 

 

 

 

 

 

 

 

 

 

displays the measured value. Do not allow vibration or

 

 

end

 

 

 

drafts to affect the balance.

 

 

 

 

 

 

 

Step

7

The balance displays

end .

 

 

Remove

 

 

 

 

Remove the weight from the pan.

 

 

 

 

Step

8

The

balance displays

glp

and outputs the

 

 

 

glp

 

 

 

"calibration test report" using the RS-232C interface

GLP output

 

 

 

 

 

 

 

or stores the calibration test data in memory. Refer to

 

 

 

end

 

 

 

"11-2. GLP Report" of the function table for details.

 

 

 

 

 

 

 

 

 

 

 

Step

9 The balance will automatically return to the weighing mode.

 

 

0 .000

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

22

 

 

 

 

 

 

g

k

g

g

Page 24
Image 24
A&D MC-10K, MC-30K manual Calibration Test Using an External Weight, Print key...... To store the new weight value. Even if

MC-10K, MC-30K specifications

The A&D MC-30K and MC-10K are highly regarded weighing scales that cater to diverse industrial and commercial applications. Renowned for their accuracy, durability, and advanced technological features, these models are ideal for users who require precision in measurement.

The MC-30K model offers a capacity of up to 30 kilograms, while the MC-10K is designed for loads of up to 10 kilograms. Both models are equipped with high-resolution load cells that provide outstanding accuracy and stability in measurements. The readability of 1 gram for the MC-10K and 5 grams for the MC-30K ensures that even the smallest variations in weight are captured effectively, making them suitable for activities ranging from laboratory analysis to quality control in manufacturing processes.

One of the standout features of these scales is their easy-to-read LCD display, which allows users to view measurements clearly, even in low-light conditions. The display is designed to provide quick and efficient readouts, minimizing wait times during the weighing process.

Both the MC-30K and MC-10K models incorporate advanced technology, including automatic calibration and tare functions, which enhance usability and accuracy. The automatic calibration feature ensures that the scales maintain their precision over time, while the tare function allows users to weigh items in containers without needing to account for the container's weight separately.

Durability is another major characteristic of these scales. Built with robust materials, they are designed to withstand the rigors of daily use in busy environments. Additionally, their compact design ensures that they occupy minimal space, making them suitable for various settings, including laboratories, warehouses, and retail outlets.

Convenience is amplified with the inclusion of RS-232 interface options for seamless connectivity to computers or printers, enabling easy data collection and reporting. The power-saving features of both models, including auto-off functionality, contribute to longer battery life, reducing the need for frequent replacements.

In summary, the A&D MC-30K and MC-10K scales are exceptional tools for anyone needing reliable and precise weighing solutions. With their advanced features, user-friendly design, and durable construction, they stand out as top choices for industrial and commercial weighing applications.