8-6. Correcting the Internal Mass Value

The balance can correct the internal mass value within the range shown below. This function corrects the internal mass value to conform to an external weight. The corrected mass value is maintained in non-volatile memory even if the AC adapter is removed. The

internal mass value is corrected as follows:

Model

Target

Range

10 kg

MC-10K

2 kg

±0.500 g

10000.00 g

MC-30K

10 kg

±5.00 g

The same mass

Correction the internal mass by +0.30 g at 10 kg. Calibrate with the corrected internal mass.

10 kg

10000.30 g

Corrected external weight

Step 1 Calibrate the balance using the internal mass. (one-touch calibration). Place the external weight and find out the correction value.

Example for correcting the weight value by +0.30 g in 10 kg, when using MC-30K. If correcting the weight value by +0.30 g in 5 kg, the weight changed into 10 kg, the correction value is +0.60 g.

Step 2 Press the

ON:OFF

key to turn off the display.

Step 3 While pressing and holding the PRINT the SAMPLE key, press the ON:OFF balance displays p5 .

key and key. The

With these keys Press the key held down

p5

Step 4 Press the PRINT key. Then the balance displays the function switches. Set the function table switch and internal mass correction switch to "1" as shown above using the following keys. SAMPLE key...To select the switch to change the value.

The selected digit blinks.

RE-ZEROkey ..To change the parameter of the switch selected. Function table switch

Correction switch of internal mass

a - 01111

a - 1 x x x 1

Step

5

Press the

PRINT

 

key to store the new setting.

 

 

The balance returns to the weighing mode.

Step

6

Press and hold the

 

 

key to enter the function

 

SAMPLE

 

 

table and release the key when

 

is displayed.

 

 

ba5fnc

0 .0 g

Press and hold

ba5fnc

Step 7 Press the SAMPLE key several times until C5 in displayed, then release the key.

Step 8 Press the PRINT key to enter the procedure for correcting the internal mass value.

is

C5 in

Step 9 Correct the internal mass value using the following keys.

0 .30 g

RE-ZEROkey..To select the value. (-500 digits appear

 

PRINT

 

after +500 digits)

key

.......

To store the new value and display the

 

next menu item of the function table.

23

Page 25
Image 25
A&D MC-30K, MC-10K manual Correcting the Internal Mass Value

MC-10K, MC-30K specifications

The A&D MC-30K and MC-10K are highly regarded weighing scales that cater to diverse industrial and commercial applications. Renowned for their accuracy, durability, and advanced technological features, these models are ideal for users who require precision in measurement.

The MC-30K model offers a capacity of up to 30 kilograms, while the MC-10K is designed for loads of up to 10 kilograms. Both models are equipped with high-resolution load cells that provide outstanding accuracy and stability in measurements. The readability of 1 gram for the MC-10K and 5 grams for the MC-30K ensures that even the smallest variations in weight are captured effectively, making them suitable for activities ranging from laboratory analysis to quality control in manufacturing processes.

One of the standout features of these scales is their easy-to-read LCD display, which allows users to view measurements clearly, even in low-light conditions. The display is designed to provide quick and efficient readouts, minimizing wait times during the weighing process.

Both the MC-30K and MC-10K models incorporate advanced technology, including automatic calibration and tare functions, which enhance usability and accuracy. The automatic calibration feature ensures that the scales maintain their precision over time, while the tare function allows users to weigh items in containers without needing to account for the container's weight separately.

Durability is another major characteristic of these scales. Built with robust materials, they are designed to withstand the rigors of daily use in busy environments. Additionally, their compact design ensures that they occupy minimal space, making them suitable for various settings, including laboratories, warehouses, and retail outlets.

Convenience is amplified with the inclusion of RS-232 interface options for seamless connectivity to computers or printers, enabling easy data collection and reporting. The power-saving features of both models, including auto-off functionality, contribute to longer battery life, reducing the need for frequent replacements.

In summary, the A&D MC-30K and MC-10K scales are exceptional tools for anyone needing reliable and precise weighing solutions. With their advanced features, user-friendly design, and durable construction, they stand out as top choices for industrial and commercial weighing applications.