34

Interval memory mode

The weighing data is periodically stored in memory.
Example For periodical weighing without a personal computer command and
to output all of the data, to a computer, at one time.
Time and date can be added with "
Time/Date output (
5-td
)"
.
Required setting dout prt 3 Stream mode / Interval memory mode
dout data 2 Data memory function is used
dout int Interval time
Optional setting dout 5-td1, 2, or 3 Adds the time and date.
10-6. Description of the Item "Data Format"

A&D standard format 5if type 0

This format is used when the peripheral equipment can receive the A&D format.
If an AD-8121B is used, set the printer to MODE 1 or 2.
This format consists of fifteen or sixteen characters excluding the terminator.
With the MC-10K, when the numerical data excluding the decimal point exceeds eight
characters, the data format will be sixteen characters long.
A header of two characters indicates the balance condition.
The polarity sign is placed before the data with the leading zeros. If the data is zero, the plus
sign is applied.
The unit, consisting of three characters, follows the data.
(Data example exceeding 8 characters)

DP (Dump print) format 5if type 1

This format is used when the peripheral equipment can not receive the A&D format.
If an AD-8121B is used, set the printer to MODE 3.
This format consists of sixteen characters excluding the terminator.
A header of two characters indicates the balance condition. No overload header is used.
The polarity sign is placed before the data, with spaces in place of leading zeros, if the data
is not zero or overloaded.
The unit, consisting of three characters, follows the data.
CRLF
ST,+10000.000 g
ST
US
OL
QT

Unit

DataHeader Ter min at or

Stable header

Unstable header

Overload header

CRLF
ST,+0000.127 g

Unit

DataHeader Terminator

Stable header (Counting mode)

WT +10000.0 00 g
CRLF
T
WT
US
QT
TerminatorHeader Data Unit
Stable header
Unstable header
Stable header (Counting mode)