16-2.

Error Codes

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Display

Error code

 

Description

 

 

 

 

 

 

 

 

 

 

 

 

Stability error

 

 

 

The balance can not stabilize due to an environmental

 

EC,E11

 

problem. Check around the pan. Refer to "3. Precautions".

 

 

Prevent vibration, drafts, temperature changes, static

 

 

 

 

 

 

electricity and magnetic fields, from influencing the balance.

 

 

 

To return to the weighing mode, press the

 

key.

 

 

 

CAL

 

 

 

Out of the setting range

 

 

 

The data to be stored is out of the setting range.

 

 

 

Internal mass error

 

EC,E16

 

Applying the internal mass does not yield a change in the

 

 

weight value as specified.

 

 

 

Confirm that there is nothing on the pan and perform the

 

 

 

weighing operation from the beginning again.

 

 

 

Internal mass error

 

EC,E17

 

The internal mass application mechanism does not function

 

 

properly. Perform the weighing operation from the beginning

 

 

 

 

 

 

again.

 

 

 

Calibration weight error

 

EC,E20

 

The calibration weight is too heavy. Confirm the calibration

 

 

weight value. Press the

 

key to return to the weighing

 

CAL

 

 

 

 

 

 

mode.

 

 

 

Calibration weight error

 

EC,E21

 

The calibration weight is too light. Confirm the calibration

 

 

weight value. Press the

 

key to return to the weighing

 

CAL

 

 

 

 

 

 

mode.

 

 

 

Overload error

 

 

 

A sample beyond the balance weighing capacity has been

 

 

 

placed on the pan. Remove the sample from the pan.

 

 

 

Weighing pan Error

 

 

 

The weight value is too light.

 

 

 

Confirm that the weighing pan is properly installed and

 

 

 

calibrate the balance.

 

 

 

Sample mass error

 

 

 

The balance can not store the sample for the counting mode

 

 

 

or for the percent mode because it is too light.

 

 

 

Use a larger sample.

 

 

 

Unit mass error

 

 

 

The sample unit mass for the counting mode is too light.

 

 

 

Storing and using it for counting will cause a counting error.

 

 

 

Add samples to reach the specified number and press the

 

 

 

 

key. Pressing the

 

 

key without adding samples

 

 

 

PRINT

 

PRINT

 

 

 

will shift the balance to the counting mode. But, to acquire

 

 

 

accurate weighing, be sure to add samples.

 

 

 

54

 

 

 

 

 

 

Page 56
Image 56
A&D MC-10K, MC-30K manual 16-2

MC-10K, MC-30K specifications

The A&D MC-30K and MC-10K are highly regarded weighing scales that cater to diverse industrial and commercial applications. Renowned for their accuracy, durability, and advanced technological features, these models are ideal for users who require precision in measurement.

The MC-30K model offers a capacity of up to 30 kilograms, while the MC-10K is designed for loads of up to 10 kilograms. Both models are equipped with high-resolution load cells that provide outstanding accuracy and stability in measurements. The readability of 1 gram for the MC-10K and 5 grams for the MC-30K ensures that even the smallest variations in weight are captured effectively, making them suitable for activities ranging from laboratory analysis to quality control in manufacturing processes.

One of the standout features of these scales is their easy-to-read LCD display, which allows users to view measurements clearly, even in low-light conditions. The display is designed to provide quick and efficient readouts, minimizing wait times during the weighing process.

Both the MC-30K and MC-10K models incorporate advanced technology, including automatic calibration and tare functions, which enhance usability and accuracy. The automatic calibration feature ensures that the scales maintain their precision over time, while the tare function allows users to weigh items in containers without needing to account for the container's weight separately.

Durability is another major characteristic of these scales. Built with robust materials, they are designed to withstand the rigors of daily use in busy environments. Additionally, their compact design ensures that they occupy minimal space, making them suitable for various settings, including laboratories, warehouses, and retail outlets.

Convenience is amplified with the inclusion of RS-232 interface options for seamless connectivity to computers or printers, enabling easy data collection and reporting. The power-saving features of both models, including auto-off functionality, contribute to longer battery life, reducing the need for frequent replacements.

In summary, the A&D MC-30K and MC-10K scales are exceptional tools for anyone needing reliable and precise weighing solutions. With their advanced features, user-friendly design, and durable construction, they stand out as top choices for industrial and commercial weighing applications.