46
13. Commands

13-1. Command List

Note A command has a terminator added, that is specified using "5if Crlf" of the
function table, and is sent to the balance.
Commands to query weighing data
C Cancels the S or SIR command.
Q Requests the weighing data immediately.
S Requests the weighing data when stabilized.
SI Requests the weighing data immediately.
SIR Requests the weighing data continuously.
Commands to control the balance
?CN Requests the upper/lower limit value code number of the selected value.
?HI Requests the upper limit value.
?ID Requests the identification number.
?LO Requests the lower limit value.
?MA Outputs all weighing data in memory.
?MQnnn Outputs data with the data number nnn. nnn: Three digits
?MX Outputs the number of data in memory (the last data number)
?PN Request the tare number of the selected value.
?PT Request the tare value.
?SN Request the serial number of the balance.
?TN Request the model name of the balance.
?UN Requests the unit mass numbers in memory.
?UW Requests the unit mass value.
CAL Same as the CAL key.
CN:mm Recalls the upper/lower limit value in memory. mm: 01 to 20.
HI:∗∗∗∗∗∗.~~g
Sets the upper limit values. ~ is space mark.
Example: the upper limit value is 2000.0 g.
Command: HI:+002000.0~~g
ID:∗∗∗∗∗∗∗ Sets identification number.
LO:∗∗∗∗∗∗.~~g
Sets the lower limit values. ~ is space mark.
Example: the lower limit value is 1000.0 g.
Command :LO:+001000.0~~g
MCL Deletes all data in memory.
MD:nnn Deletes data with the data number nnn. nnn: Three digits.
OFF Turns the display off.
ON Turns the display on.
P Same as the ON:OFF key
PN:mm Recalls the tare value in memory. mm: 01 to 20.
PRT Same as the PRINT key
PT:∗∗∗∗∗.~~g
Sets the tare value. ~ is space mark.
Example: the tare value is 1000.0 g.
Command :PT:+001000.0~~g