HE/001132

Page 15 of 92

 

Document No.001132, Revision 2.0

Agfa HealthCare

NodeID Livelink : 14140190

14 June, 2007

 

 

Table 2.2-7: Association Reject Reasons

Result

Response

Reason/Diag

The Dicom library refuses the association

 

SCU Device not known

 

 

 

The Dicom library refuses the association

 

Maximum number of association

 

 

processing threads exceeded

 

 

 

The Dicom library refuses the association

A799

Dicom Library is not allowed to accept

 

 

C-STORE commands from this device

 

 

and therefore refuses the association.

The Dicom library refuses the association

A0102

SOP Class is not found back in the

 

 

configuration

The Dicom library refuses the association

A702

The Dicom library is unable to create the

 

 

Dicom media file due to the fact that the

 

 

disk is full

The Dicom library refuses the association

A703

The Dicom library is unable to create the

 

 

Dicom media file due to resource

 

 

problems other than ‘disk is full’.

The Dicom library refuses the association

A703

The Dicom library is unable to acquire

 

 

the complete C-STORE request due to

 

 

network problems.

The Dicom library refuses the association

A701

The Dicom library is unable to acquire

 

 

the complete C-STORE request due to

 

 

network problems (time-out while

 

 

reading data from socket) 2.

2.2.1.3.1.2Proposed Presentation Contexts

The DICOM Store AE is capable of proposing the Presentation Contexts shown in the following table:

2The time-out value is hard-coded in the Dicom library as being 60 seconds.

Page 15
Image 15
AGFA HE/001132 manual Proposed Presentation Contexts, Association Reject Reasons, Result Response Reason/Diag