HE/001132 | Document No. 001132, Revision 2.0 | Page 24 of 92 |
Agfa HealthCare |
| |
NodeID Livelink : 14140190 | 14 June, 2007 | |
|
|
|
(0070,0001) | Graphic Annotation Sequence |
|
(0070,0022) | >> Graphic Data | [values] |
(0070,0023) | >> Graphic Type | POLYLINE |
(0070,0024) | >> Graphic Filled | - |
2.2.1.3.1.3.1.6.9Calibration ruler
The calibration ruler is displayed at the side of the image and is the result of a calibration action of the operator.
It consists out of a ruler (constructed using several lines) and some text.
Graphic Layer Module
|
| (0070,0060) |
| Graphic Layer Sequence |
|
|
| (0070,0002) |
| >Graphic Layer | ID |
|
| (0070,0068) |
| >Graphic Layer Description | Calibration Ruler |
Graphic Annotation Module |
| ||||
|
| (0070,0001) |
| Graphic Annotation Sequence |
|
|
| (0070,0002) |
| > Graphic Layer | [ID of the corresponding layer] |
|
| (0070,0008) |
| > Text Object Sequence |
|
|
| (0070,0003) |
| >> Bounding box annotation units | PIXEL |
| (0070,0004) |
| >> Anchor Point Annotation Units | - | |
| (0070,0006) |
| >> Unformatted Text Value | [value] | |
| (0070,0010) |
| >> Bounding Box Top Left Hand Corner | [value] | |
| (0070,0011) |
| >> Bounding Box Top Right Hand Corner | [value] | |
| (0070,0012) |
| >> Bounding Box Text Horizontal Justification | LEFT | |
| (0070,0014) |
| >> Anchor Point | - | |
| (0070,0015) |
| >> Anchor Point Visibility | - | |
|
| (0070,0009) |
| > Graphic Object Sequence |
|
|
| First ruler line |
|
|
|
| (0070,0005) |
| >> Graphic Annotation Units | PIXEL | |
| (0070,0020) |
| >> Graphic Dimensions | 2 | |
| (0070,0021) |
| >> Number of Graphic Points | 2 | |
| (0070,0022) |
| >> Graphic Data | [values] | |
| (0070,0023) |
| >> Graphic Type | CIRCLE | |
| (0070,0024) |
| >> Graphic Filled | N | |
|
| … |
|
|
|
2.2.1.3.1.3.1.6.10Angle
An angle measurement is a combination of 2 lines and 2 short lines, with 2 angle texts