HE/001132

Document No. 001132, Revision 2.0

Page 24 of 92

Agfa HealthCare

 

NodeID Livelink : 14140190

14 June, 2007

 

 

 

(0070,0001)

Graphic Annotation Sequence

 

(0070,0022)

>> Graphic Data

[values]

(0070,0023)

>> Graphic Type

POLYLINE

(0070,0024)

>> Graphic Filled

-

2.2.1.3.1.3.1.6.9Calibration ruler

The calibration ruler is displayed at the side of the image and is the result of a calibration action of the operator.

It consists out of a ruler (constructed using several lines) and some text.

Graphic Layer Module

 

 

(0070,0060)

 

Graphic Layer Sequence

 

 

 

(0070,0002)

 

>Graphic Layer

ID

 

 

(0070,0068)

 

>Graphic Layer Description

Calibration Ruler

Graphic Annotation Module

 

 

 

(0070,0001)

 

Graphic Annotation Sequence

 

 

 

(0070,0002)

 

> Graphic Layer

[ID of the corresponding layer]

 

 

(0070,0008)

 

> Text Object Sequence

 

 

 

(0070,0003)

 

>> Bounding box annotation units

PIXEL

 

(0070,0004)

 

>> Anchor Point Annotation Units

-

 

(0070,0006)

 

>> Unformatted Text Value

[value]

 

(0070,0010)

 

>> Bounding Box Top Left Hand Corner

[value]

 

(0070,0011)

 

>> Bounding Box Top Right Hand Corner

[value]

 

(0070,0012)

 

>> Bounding Box Text Horizontal Justification

LEFT

 

(0070,0014)

 

>> Anchor Point

-

 

(0070,0015)

 

>> Anchor Point Visibility

-

 

 

(0070,0009)

 

> Graphic Object Sequence

 

 

 

First ruler line

 

 

 

 

(0070,0005)

 

>> Graphic Annotation Units

PIXEL

 

(0070,0020)

 

>> Graphic Dimensions

2

 

(0070,0021)

 

>> Number of Graphic Points

2

 

(0070,0022)

 

>> Graphic Data

[values]

 

(0070,0023)

 

>> Graphic Type

CIRCLE

 

(0070,0024)

 

>> Graphic Filled

N

 

 

 

 

 

2.2.1.3.1.3.1.6.10Angle

An angle measurement is a combination of 2 lines and 2 short lines, with 2 angle texts

Page 24
Image 24
AGFA HE/001132 manual Calibration ruler