HE/001132 | Page 69 of 92 |
| Document No.001132, Revision 2.0 |
Agfa HealthCare | NodeID Livelink : 14140190 |
14 June, 2007 | |
|
|
Exceptions exist for attributes which are necessary to correctly import or interpret the image. Therefore the following values are kept or replaced by generated values:
Table 3.2-2: Necessary Values when performing an Anonymous Export.
Attribute | Tag | Set to … |
Patient’s Name | (0010,0010) | “Anonymous” |
Study Instance UID | (0020,000D) | Keep value |
Modality | (0008,0060) | Keep value |
Series Instance UID | (0020,000E) | Keep value |
Laterality | (0020,0060) | Keep value |
Series description | (0008,103E) | Keep value |
Protocol name | (0018,1030) | Keep value |
Body Part Examined | (0018,0015) | Keep value |
Smallest Pixel Value in Series | (0028,0108) | Keep value |
Largest Pixel Value in Series | (0028,0109) | Keep value |
Presentation Intent Type | (0008,0068) | Keep value |
SOP Instance UID | (0008,0018) | Keep value |
SOP Class UID | (0008,0016) | Keep value |
Specific character set | (0008,0005) | Keep value |
Patient Orientation | (0020,0020) | Keep value |
Image type | (0008,0008) | Keep value |
Image comments | (0020,4000) | Keep value |
Presentation LUT shape | (2050,0020) | Keep value |
3.2.1.1.1.1Media Storage Application Profile
The
3.2.1.1.1.1.1Options
The
Table
| IOD | SOP Class UID |
|
| Transfer Syntax | Transfer Syntax UID |
|
| CR Image Storage | 1.2.840.10008.5.1.4.1.1.1 |
|
| Explicit VR Little Endian | 1.2.840.10008.1.2.1 |
|
|
|
|
|
|
|
|
|
| Digital | 1.2.840.10008.5.1.4.1.1.1.1 |
|
| Explicit VR Little Endian | 1.2.840.10008.1.2.1 |
|
| Storage – For Presentation |
|
|
| |||
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|