AGFA HE/001132 manual File Meta Information for Implementation Class and Version

Models: HE/001132

1 93
Download 93 pages 37.83 Kb
Page 67
Image 67

HE/001132

Page 67 of 92

 

Document No.001132, Revision 2.0

Agfa HealthCare

NodeID Livelink : 14140190

14 June, 2007

 

 

3MEDIA INTERCHANGE

"Workstation" is able to create or read DICOM Interchange media. The related capabilities are described in the following sections.

3.1Implementation Model

Export To CD-R

Offline-Media Application

Entity

Export

CD-R

Storage Medium

Figure 3.1-1: Application Data Flow Diagram for Media Storage

-The Offline-Media Application Entity exports DICOM images and Presentation States to a CD-R Storage medium. It is associated with the local real-world activity “Export Images”. “Export Images” is performed upon user request.

3.1.2Functional Definition of AEs

3.1.3Sequencing of Real World Activities

At least one image or presentation state must exist and be selected before the Offline-Media Application Entity can be invoked. The Offline-Media Application Entity is invoked through the local real-world activity “Export Images”.

The operator must insert a new (blank) CD-R media before invocation of the Offline-Media Application Entity. If no CD-R is inserted, the Offline-Media Application Entity will prompt for a media to be inserted before starting to write to the CD-R device. The export job can be canceled from the job queue by clicking “Cancel” on this prompt.

Processed Images are exported in standard DICOM format. A Dicom Viewer is burned with the images on the CD, to view these images later on. The images can be read by any Dicom Compatible application. Export to a hard disk is NOT supported by "Workstation".

3.1.4File Meta Information for Implementation Class and Version

The implementation information written to the File Meta Header in each file is:

Page 67
Image 67
AGFA HE/001132 File Meta Information for Implementation Class and Version, Application Data Flow Diagram for Media Storage