HE/001132

Document No. 001132, Revision 2.0

Page 64 of 92

Agfa HealthCare

 

NodeID Livelink : 14140190

14 June, 2007

 

 

 

on forward the images to the PACS. A DICOM Store queue can only be rerouted to another DICOM Store queue with

-The same output type configured (this includes the SOP class to be used)

-The same GSPS support

If the new destination is configured for Storage Commitment, the storage commit request will be sent to the new destination when processing the job.

By means of a SENT-flag, an image that has already been successfully sent to a specific archive cannot be sent to that archive twice. The SENT-flag can be applied to all DICOM store destinations, including both the archive and the other softcopy destinations (e.g. viewing stations).

2.4.1.4.2Job description

One archive job can contain multiple images. All these images and their GSPS’s will be sent to the archive through a single association.

The user can perform the following actions on existing jobs: “Delete job” and “Expedite job”

When an image in the archive job does not reach its destination, or when the storage commit replies “time out”, the job is FAILED. .

2.4.1.5DICOM Print AE

There is only one entry per print job (with possible multiple sheets) in the queue.

All queue actions are done on one print job: retry, reprint, expedite, …

2.4.2Parameters

The specification of important operational parameters, and if configurable, their default value and range, are specified in the table below.

The parameters that apply to all Application Entities are specified in the “General Parameters” section. Those specific to particular Applications are specified in separate sections specific to each AE.

Table 2.4-5: Configurable Parameters for "Workstation".

Parameter

 

Configurable

Default value

 

 

 

(yes/no)

 

 

General Parameters

 

 

 

 

Max PDU Receive Size

 

No

65542

 

 

 

 

 

 

Max PDU Send Size (larger PDUs will never be

 

 

 

 

sent, even if the receiver supports a larger Max PDU

 

 

 

 

Receive Size. If the receiver supports a smaller Max

 

 

 

 

PDU Receive Size then the Max PDU Send Size will

 

No

65542

 

be reduced accordingly for the duration of the

 

 

Association. Max PDU Receive Size information is

 

 

 

 

exchanged during DICOM Association Negotiation in

 

 

 

 

the Maximum Length Sub-Item of the A-

 

 

 

 

ASSOCIATION-RQ and A-ASSOCIATE-AC)

 

 

 

 

Page 64
Image 64
AGFA HE/001132 manual Job description, Dicom Print AE, Configurable Parameters for Workstation