Agilent Technologies 53150A, 53152A Power Correction Theory of Operation, Using Power Correction

Models: 53151A 53150A 53152A

1 142
Download 142 pages 17.82 Kb
Page 85
Image 85
Power Correction Theory of Operation

Chapter 2 Operating Your Frequency Counter

Using Power Correction

Power Correction Theory of Operation

When the Counter interpolates between data points to determine the amount of correction to apply to the current measurement, it computes the correction based on a straight line plotted between the frequency values in the two closest data points. Therefore, a graph of a power-correction profile would show a “curve” that consists of two to nine straight-line segments, rather than a true curve, as shown below.

P

o

w

e

r

 

F r e q u e n c y

 

 

NOTE

When the frequency value of a signal is beyond either end of the curve

 

specified by the values in the currently selected profile’s data points,

 

the Counter computes a correction by extending the straight line defined

 

in the last two data points in the profile.

 

As the graph above shows, the Counter never computes power-correction

 

 

values for loss above the zero axis. Conversely, corrections are never

 

computed for gain below the zero axis. When the Counter interpolates for

 

frequencies that are above or below the range of frequencies specified in

 

the currently selected profile, it never computes a value that would fall on

 

the opposite side of the zero axis from the closest specified frequency.

 

In other words, if there are two or more data points that contain loss

 

values, the Counter never computes or applies a correction that would be

 

indicative of gain. Conversely, if there are two or more data points that

 

contain gain values (negative loss values), the Counter never applies a

 

correction that would be indicative of loss.

2

Operating Guide

2-31

Page 85
Image 85
Agilent Technologies 53150A, 53152A, 53151A manual Power Correction Theory of Operation, Using Power Correction