profile are sorted by frequency and stored in that order. Therefore,

2

Chapter 2 Operating Your Frequency Counter

Using Power Correction

10To add or adjust the values in another data point, press the left-arrow key repeatedly until “PWR CORR” is re-displayed, and repeat steps 2 through 10. To edit data points within another power-correction profile, press the left-arrow key repeatedly until “PWR CORR” is re-displayed, and repeat steps 1 through 10.

NOTE

When a data point is entered in a profile, all of the data points in that

 

profile are sorted by frequency and stored in that order. Therefore,

 

the data points in a profile are always in order from the lowest frequency

 

to the highest when you access a correction profile. The first data point

 

displayed is always the data point that contains the lowest frequency value.

 

 

11If you are done entering data-point values, press the Enter key to accept your changes, save the data-point values, and return to the measurement display.

Power Correction is now enabled using the profile number 3.

 

Ch 2

 

Freq

 

dBm

 

Pwr

NOTE

You can include your selection of a power-correction profile in any of the

 

nine sets of user settings stored in non-volatile memory (see page 2-46) by

 

enabling the profile and then saving the set. If you save set 0 (zero) while

 

Power-Correction is enabled, the currently-selected profile will be in effect

 

whenever you turn on the Counter.

The data in the saved user settings and in all of the power-correction profiles is stored in non-volatile memory. If the Counter ever requires repair, and the main circuit-board assembly is replaced, these stored values will be lost. Therefore, to protect these values and ensure they are available for future use, you should keep an external record of them.

2-38

Operating Guide