Agilent Technologies 53151A, 53152A, 53150A Power Correction Example Editing Data Point Values

Models: 53151A 53150A 53152A

1 142
Download 142 pages 17.82 Kb
Page 89
Image 89
Power Correction Example: Editing Data Point Values

Chapter 2 Operating Your Frequency Counter

Using Power Correction

Power Correction Example: Editing Data Point Values

1Select a power-correction profile using the up- and/or down-arrow keys (as shown in steps 1 through 4 in the previous example). For this example, choose profile number 3.

2Press the right-arrow key.

The display changes to show the loss and frequency values stored in the data point that contains the lowest frequency value in the current profile. The left and right pointers that are before and after the word “LOSS” are flashing.

dB

The ten small, single-line annunciators that are below the analog power display are used to indicate which of the ten data points is currently displayed. In the example shown above, only the first annunciator is activated, which indicates that the values displayed are for data point 1, the lowest data point.

2

Operating Guide

2-35

Page 89
Image 89