Replacing the Batteries

Appendix C Using the Battery Option

Replacing the Batteries

C

Main ~ Power

Reference 10 MHz

Auxillary

 

 

In

 

 

or

 

 

Out

 

 

11 TO 18 VDC

 

 

EXT DC

RS-232

Made in U.S.A.

with domestic and foreign content

ISM 1-A

OPTIONS

Figure C-2. External DC Power Socket 001 Oven Time Base

Removing the Batteries 002 Battery

WARNING:

To avold electric shock, do not remove covers.

No user-serviceable parts inside.

Refer all servicing to qualified personnel.

This unit must be earth grounded.

AC POWER

FUSE

100

– 130 VAC, 50/60/400 Hz 75 VA

1.0 A T

220

– 240 VAC, 50/60 Hz 75 VA

250 V

Main AC Power Indicator

External DC Input Connector

Figure C-2. External DC Power Socket

Replacing the Batteries

Counters equipped with the Battery option use sealed lead-acid VHS camcorder batteries. You can obtain additional batteries of this type from Agilent (Agilent P/N 53150-80010) and from other suppliers who carry test equipment and/or video camera accessories.

Removing the Batteries

To remove the batteries, use the following procedure:

1Turn off the Counter, and disconnect all external cables (including the power cord).

2Turn the two thumb screws that secure the battery sled to the Counter’s back panel counterclockwise until both thumb screws are completely out of the threaded holes in the back panel (see Figure C-3).