Chapter 3 Specifications

Timebase

Frequency: 10 MHz

Output: 10 MHz sine wave, 1 Vrms into 50 Ω

External Timebase Input: 1, 2, 5, 10 MHz;

1 to 5 Vrms into 50 Ω

Connector: BNC female located on rear panel

Internal Timebase Stability

 

TCXO

Oven

 

(Standard)

(Option 001)

Aging Rate

<5 x 1010

Per Day

Per Month

<1 x107

<1.5 x 108

Short Term

<1 x 109

<2 x 1010

(1 sec. avg. time)

Line Variation

<1 x 107

<1 x 1010

(±10%)

Warm-up

<1 x108

 

 

within 5 min. after

 

 

turn-on at 25°C

Temperature

<1 x 106

<1 x 108

Stability (0–55°C)

General Information

Save and Recall: Up to 9 complete instrument setups may be saved and later recalled. These setups are retained when power is removed.

Sample Rate: User-selectable Fast (nominally 20 ms between readings), Medium (nominally 250 ms between readings), Slow (nominally 1 s between readings) and Hold.

Self Test: Internal memory and count circuitry automatically tested at startup, via menu selection, or remotely. Error messages displayed to indicate failed tests.

Size: 213 mm W x 88.5 mm H x 300 mm D

Operating temperature: 0–55°C With battery option: 0–40°C

Weight: 4 kg without battery option, 6.4 kg with battery option

Warranty: 1 year

Programming: GPIB (IEEE-488.1-1987, IEEE 488.2-1987) or RS-232C

Language: SCPI-1992.0 (Standard Commands for Programmable Instruments)

RS-232C Rates: User-selectable 2400 to 19200 baud

Power Supply

ac: 90–132 Vac; 47.5–66 Hz or 360–440 Hz 216–264 Vac; 47.5–66 Hz

line selection: automatic

power requirements: 75 VA max. (25 W typ.) dc: (Option 002 only): 11–18 Vdc; 2A max. Battery (Option 002):

Type: VHS camcorder, lead acid (2 each)

Charge Time: 8 hours in unit

Capacity: 2.5 hours min. at 25°C Math Functions:

Offset: Last reading and/or entered offset to reading for either power or frequency Averaging: 1 to 99 measurement running average

Cable Loss Compensation: Offsets power reading via linear interpolation of user-entered attenuations with up to 9 independent frequency points.

Display: Backlit LCD. Backlight can be turned on or off via front panel control.

Sleep Mode (Option 002 only): Automatically activated if no input is present for 5 minutes.

Safety: Designed in compliance with IEC-1010, CAN/CSA 1010.1

EMC: Designed in compliance with IEC-11, EN50082-1, IEC801-2, -3, -4

Accessories Supplied

Operating, programming, and service manuals and ac power cord.

Accessories Available

Spare Battery 53150-80010 dc Power Input Cable 53150-60214

Options

Opt 001 Oven Timebase

Opt 002 Battery and dc input

Opt 1BP Mil-Std-45662A Calibration with data Opt W30 Three Years of Return Repair Service Opt W50 Five Years of Return Repair Service Opt 1CM Rack Mount Kit

3

Operating Guide

3-7

Page 107
Image 107
Agilent Technologies 53151A, 53152A, 53150A manual Accessories Supplied Accessories Available

53151A, 53150A, 53152A specifications

Agilent Technologies, renowned for its cutting-edge electronic measurement equipment, produced a range of high-performance frequency counters, among which the 53150A, 53151A, and 53152A models stand out. These versatile instruments are designed to cater to various measurement applications in telecommunications, aerospace, and research and development.

The Agilent 53150A is a high-performance frequency counter that operates on a wide frequency range, capable of measuring frequencies up to 2.7 GHz. This model features a 10 Hz to 2.7 GHz measurement capability with a time interval resolution of 100 ps, making it ideal for accurately capturing fast signals. Additionally, it offers a built-in statistical analysis feature that enables users to perform precise measurements, minimizing uncertainty in data collection.

Equally impressive, the Agilent 53151A enhances capabilities further by introducing advanced measurement functions. This model supports frequency and time interval measurements up to 3.5 GHz and includes an additional mode for measuring CW frequency and period. The enhanced resolution of 50 ps allows users to capture even finer variations in measured signals. The 53151A also features a user-friendly graphical display, simplifying data visualization and interpretation.

The Agilent 53152A pushes performance to new heights with frequency measurements extending up to 7 GHz. This model incorporates a high-speed external timebase for increased accuracy and uses advanced signal processing technology to ensure reliable measurements, even in noisy environments. One of the standout features of the 53152A is its capability to interface with a variety of devices, enabling remote measurements and integration into automated testing environments.

All three models employ state-of-the-art technology and boast characteristics that enhance their usability. They feature multi-channel capability, allowing simultaneous measurements, which is critical for complex applications. Their robust design and compact form factor make them suitable for both laboratory and field use, ensuring versatility in different working environments.

In summary, Agilent Technologies’ 53150A, 53151A, and 53152A frequency counters combine exceptional measurement accuracy, a wide frequency range, and advanced functionalities tailored for professionals across various scientific and engineering disciplines. Whether measuring signal frequency, time intervals, or conducting statistical analyses, these models are engineered to meet the demanding requirements of today’s high-tech world.