Agilent Technologies 669xA, 665xA, 664xA, 667xA, 668xA manual Tst?, Meaning and Type, Description

Models: 668xA 669xA 667xA 664xA 665xA

1 70
Download 70 pages 51.66 Kb
Page 33
Image 33

*TRG

Meaning and Type

Trigger Device Trigger

Description

This command generates a trigger when the trigger subsystem has BUS selected as its source. The command has the same affect as the Group Execute Trigger (<GET>) command.

Command Syntax

*TRG

Parameters

(None)

Query Syntax

(None)

Related Commands

ABOR CURR:TRIG INIT TRIG[:IMM] VOLT:TRIG

 

<GET>

*TST?

Meaning and Type

Test Device Test

Description

This query causes the power supply to do a self-test and report any errors (see "Selftest Error Messages" in "Chapter 3 - Turn-On Checkout" of the power supply Operating Guide).

Query Syntax

*TST?

 

Returned Parameters

<NR1>

 

 

0

Indicates power supply passed self-test.

Related Commands

Nonzero

Indicates an error code.

(None)

 

*WAI

Meaning and Type

Wait to Continue Device Status

Description

This command instructs the power supply not to process any further commands until all pending operations are completed. "Pending operations" are as defined under the *OPC command. *WAI can be aborted only by sending the power supply a GPIB DCL (Device Clear) command.

Command Syntax

*WAI

Parameters

(None)

Query Syntax

(None)

Related Commands

*OPC *OPC?

Language Dictionary 33

Page 33
Image 33
Agilent Technologies 669xA, 665xA, 664xA, 667xA, 668xA manual Tst?, Meaning and Type, Description