*RST

Meaning and Type

Reset Device State

Description

This command resets the power supply to a factory-defined state as defined below. *RST also forces an ABORt command.

 

Command

State

 

CAL:STAT OFF

OUTP[:STAT] OFF

CURR[:LEV][:IMM] *

OUTP:PROT:DEL *

CURR[:LEV]:TRIG *

OUTP:REL[:STAT] OFF

CURR:PROT:STAT OFF

OUTP:REL:POL NORM

DIG:DATA 0

TRIG:SOUR BUS

DISP[:WIND]:STAT ON

VOLT[:LEV][:IMM] *

DISP[:WIND]:MODE NORM

VOLT[:LEV][:TRIG] *

DISP[:WIND]:TEXT

VOLT:PROT[:LEV] *

INIT:CONT OFF

* Model-dependent value. See Table 3-1.

Command Syntax

*RST

 

Parameters

(None)

 

Query Syntax

(None)

 

Related Commands

*PSC

*SAV

*SAV

Meaning and Type

SAVE Device State

Description

This command stores the present state of the power supply to the specified location in memory. Up to four states can be stored. Under certain conditions (see "Turn-On Conditions" in "Chapter 5 - Front Panel Operation" of the Operating Guide), location 0 may hold the device state that is automatically recalled at power turn-on.

The following power supply parameters are stored by *SAV:

CURR[:LEV][:IMM]

OUTP[:STAT]

OUTP:REL:POL

CURR:PROT:STAT

OUTP:PROT:DEL

VOLT[:LEV][:IMM]

DIG:DATA[:VAL]

OUTP:REL[:STAT]

VOLT:PROT[:LEV]

Command Syntax

*SAV <NRf>

 

Parameters

0123

 

Example

SAV 3

 

Query Syntax

(None)

 

Related Commands

*RCL *RST

 

The power supply uses nonvolatile memory for recording register states. Programs that repeatedly use *SAV for recalling states cause frequent write cycles to the memory and can eventually exceed the maximum number of write cycles for the memory (see in the power supply Operating Guide).