Description of Subsystem Commands
Figure
Figure 3-2. Subsystem Commands Tree Diagram
ABOR
This command cancels any trigger actions presently in process. Pending trigger levels are reset equal to their corresponding immediate values. ABOR also resets the WTG bit in the Operation Condition Status register (see "Chapter 4 - Status Reporting"). If INIT:CONT ON has been programmed, the trigger subsystem initiates itself immediately after ABORt, thereby setting WTG. ABOR is executed at power turn on and upon execution of *RCL or RST.
Command Syntax | ABORt |
Parameters | (None) |
Examples | ABOR |
Query Syntax | (None) |
Related Commands | INIT *RST *TRG TRIG |
Calibration Commands
See Appendix A in the power supply Operating Guide.
34 Language Dictionary