Use, Maintenance, and Care of the Devices

Gaging Connectors

Gaging Female Type-N Connectors

NOTE

Always hold a connector gage by the gage barrel, below the dial indicator.

 

This gives the best stability, and improves measurement accuracy.

 

 

1.Select the proper gage for your connector. (Refer to Table 6-3for gage part numbers).

2.Inspect and clean the gage, gage master, and device to be gaged. Refer to “Visual Inspection” and “Cleaning Connectors” earlier in this chapter.

3.Zero the connector gage (refer to Figure 3-4):

a.While holding the gage by the barrel, and without turning the gage or the gage master, screw the gage connecting nut onto the female gage master, just until you meet resistance. Connect the nut finger-tight. Do not overtighten.

b.Use the torque wrench recommended for use with this kit to tighten the connecting nut to 135 N-cm (12 in-lb). Refer to “Connections” on page 3-12for more information.

c.Loosen the dial lock screw on the gage and rotate the gage dial so that the pointer corresponds to the correction value noted on the gage master. Do not adjust the gage dial to zero, unless the correction value on the gage master is zero.

d.Tighten the dial lock screw and remove the gage master.

e.Attach and torque the gage master to the gage once again to verify that the setting is repeatable. Remove the gage master.

4.Gage the device connector (refer to Figure 3-4):

a.While holding the gage by the barrel, and without turning the gage or the device, screw the gage connecting nut onto the device being measured, just until you meet resistance. Connect the nut finger-tight. Do not overtighten.

b.Use the torque wrench recommended for use with this kit to tighten the connecting nut to 135 N-cm (12 in-lb). Refer to “Connections” on page 3-12for more information.

c.Gently tap the barrel of the gage with your finger to settle the gage reading.

d.Read the gage indicator dial. If the needle has moved clockwise, the center conductor is protruding by an amount indicated by the black numbers. If the needle has moved counterclockwise, the center conductor is recessed by an amount indicated by the red numbers.

For maximum accuracy, measure the connector a minimum of three times and take an average of the readings. After each measurement, rotate the gage a quarter-turn to reduce measurement variations that result from the gage or the connector face not being exactly perpendicular to the center axis.

e.Compare the average reading with the observed pin depth limits in Table 2-2 on page 2-4.

3-10

85032B/E

Page 26
Image 26
Agilent Technologies 85032B/E 50 manual Gaging Female Type-N Connectors