Examples of upgrade paths

Examples of upgrade paths

85107B

(45 MHz to 50 GHz) current system

Upgrade to 85106D waveguide system to operate in V-band

8510C

8517B

83651B

New equipment needed:

85105A1 with Option 0502 millimeter- wave test set controller

83621B synthesizer (LO source)

V85104A 3 test set modules (quantity of 2)

V11644A 3 calibration kit

V11645A 3 verification kit

8510C

85105A #050

8517B

83651B

83621B

Figure 10. Upgrading an 85107B 50 GHz system to an 85106D waveguide system to operate in V-band (or W-band)

Measurement accessories

There are measurement accessories for seven device connector types: 7 mm,

3.5mm, 2.92 mm, 2.4 mm, 1.85 mm, 1 mm and Type-N. Calibration kits include standards that are required for vector error correction. Verification kits include standards used to verify system performance specifications. Test port return cables extend the ports of the test set and connect to the device under test. Agilent 85130X adapter sets convert test set ports to the same connector type (acting as a test port saver) or to a different connector type.

1.Standard 85105A already includes IF switching (Option 001)

2.Option 050 provides a 50 GHz RF source switch. This is necessary in order to operate with the existing 83651B 50 GHz source and 8517B 50 GHz test set.

3. To operate in W-band, replace all V-band components with W-band.

15