Calibration kits

Calibration kits

Error correction requires that the systematic errors in the measurement system be characterized by measuring known devices (standards) over the frequency range of interest with the process of calibration. All calibration kits contain standards used for this purpose. The standards in the 3.5 mm, 2.4 mm and Type-N calibration kits use the precision slotless connector (PSC-3.5, PSC-2.4 and PSC-N). Unless otherwise noted all coaxial calibration kits include connector gauges and a torque wrench.

Mechanical

Connector

Frequency

Description

calibration

type

range

 

kit

 

(GHz)

 

 

 

 

 

 

85050B

7 mm

0.045

- 18

Contains open and short circuits, fixed and sliding terminations.

 

 

 

 

 

85050C

7 mm

0.045

- 18

Precision kit. Contains standards for TRL calibration, including precision airline.

 

 

 

 

Also contains open and short circuits and fixed termination.

 

 

 

 

 

85050D

7 mm

0.045

- 18

Economy kit. Contains open and short circuits and precision fixed termination.

 

 

 

 

Gauges not included.

85052B

3.5 mm

0.045

- 26.5

Contains open and short circuits, fixed and sliding terminations and in-series

 

 

 

 

adapters.

 

 

 

 

 

85052C

3.5 mm

0.045

- 26.5

Precision kit. Contains standards for TRL calibration, including precision airlines.

 

 

 

 

Also contains open and short circuits, fixed terminations and in-series adapters.

 

 

 

 

Gauges not included.

 

 

 

 

 

85052D

3.5 mm

0.045

- 26.5

Economy kit. Contains open and short circuits, precision fixed termination, and in-series

 

 

 

 

adapters. Gauges not included.

 

 

 

 

 

85054B

Type-N

0.045

- 18

Contains open and short circuits, fixed and sliding terminations, in-series adapters

 

 

 

 

and 7 mm-to-Type-N adapters.

85054D

Type-N

0.045

- 18

Economy kit. Contains open and short circuits, fixed terminations, in-series

 

 

 

 

adapters and 7 mm-to-Type-N adapters. Gauges not included.

 

 

 

 

 

85056A

2.4 mm

0.045

- 50

Contains open and short circuits, fixed and sliding terminations and in-series adapters.

85056D

2.4 mm

0.045

- 50

Economy kit. Contains open and short circuits, fixed terminations and in-series

 

 

 

 

adapters. Gauges not included.

 

 

 

 

 

85056K

2.92/2.4 mm

0.045

- 50

Contains 2.4 mm open and short circuits, fixed loads and 2.92 mm adapters.

85059A

1 mm

0.045

- 110

Broadband coaxial precision calibration kit consists of 1 mm shorts, opens, fixed

 

 

 

 

loads and in-series adapters. It also includes offset-shorts covering 50 to 110 GHz.

 

 

 

 

Gauges not included.

11904S

2.92 mm

0.045

- 40

Must be used with Agilent 85056A/D 2.4 mm calibration kit. Includes four 2.92 mm-to-

 

 

 

 

2.4 mm adapters. Gauges not included.

 

 

 

 

 

X11644A

WR-90

8.2 - 12.4

P11644A

WR-62

12.4 - 18

 

 

 

 

K11644A

WR-42

18

- 26.5

 

 

 

R11644A

WR-28

26.5 - 40

Q11644A

WR-22

33

- 50

 

 

 

 

U11644A

WR-19

40

- 60

 

 

 

 

V11644A

WR-15

50

- 75

W11644A

WR-10

75

- 110

Contains standards for TRL calibration. Includes precision waveguide section, short circuit and fixed or sliding terminations. Gauges not included.

16

Page 16
Image 16