FM Receivers
90
If desired, use the meter averaging function for the SINAD indicator.
aSelect dB on the SINAD meter.
bPress the AVG key.
If desired, select the number of readings to average.
Enter the number of re adings to average.
Higher number of readings averaged will require longer to
reach the final indication.
16 Increase Amplitude (more positive) by 6 dB.
17 Increase RF Gen Freq (in 100 Hz steps) until the SINAD meter again reads
12 dB.
Record the frequency as High Freq for use later in the procedure.
18 Decrease RF Gen Freq (in 100 Hz steps) until the SINAD meter again
reads 12 dB.
Record the frequency as Low Freq for use later in the procedure.
Calculate the measurement:
19 High Freq Assigned Freq and Assigned Freq Low Freq
The minimum usable bandwidth is the smaller result.