Using the Oscilloscope
237
to the RF IN/OUT o r ANT IN connectors .
Audio In for a signal connected to the AUDIO IN connector.
Radio Int for a signal connected to the optional rear panel Radio Inter-
face connector.
Ext Mod for a signal connected to the rear panel MODULATION IN-
PUT connector.
Mic Mod for a signal connected to the MIC/ACC connector “MIC” pin.
FM Mod for the FM modulated audio from the RF Gen section.
AM Mod for the AM modulated audio from the RF Gen section.
Audio Out for the signal present at the AUDIO OUT connector
5. Select from the Scope To field’s list of choices where in t he AF Analyzer’s
circuitry the signal is routed to the Oscilloscope.
NOTE: All choices except Input are capa citiv e couple d. Use Input if the sign al bein g
measured is 1Hz.
Input to route the audio to the Oscilloscope without being processed.
Filters to route the audio to the Oscilloscope after pa ssing through Fil-
ters #1 and #2.
De-emp to route the audio to the Oscilloscope after passing through F il-
ters #1 and #2, and the De-Emphasis circuitry.
Notch to route the audio to the Oscilloscope after passing through Fil-
ters #1 and #2, the De-Emphasis circuitry, and Notch circuitry.
6. Select the SCOPE screen.
7. Select the CONTROLS field.
8. Continue the measurement by selecting a control screen from the list of
choices and referring to the following:
see "Measurements U sing the Main Contro l Fields:" on page 2 38.
see "Measurement Using the Trigger Control Fields:" on page 239.
see "Measurements Using th e Ma rk e r C o n tro l Fie ld s: " o n pag e
240.