Chapter 3 Functions and Operations
Table
Error Message | Description |
CANNOT ZERO | Attempting to perform a Power Monitor zero adjust function with an |
INPUT SIGNAL | input of greater than |
TOO HIGH |
|
POWER MONITOR | Attempting to enter Power Monitor mode with no option 5 installed. |
OPTION NOT |
|
INSTALLED |
|
LIMIT SEGMENTS | Segment limit lines are not allowed to overlap in frequency or dis- |
CANNOT | tance. |
OVERLAP |
|
EITHER STORED | Attempt to overlay a stored trace to the display from an empty loca- |
SWEEP IS NOT | tion, or with a different measurement mode or setup than the current |
COMPATIBLE, OR | display. |
NO STORED |
|
SWEEP AT THIS |
|
LOCATION |
|
VERTICAL LIMIT | Segment limit line does not support vertical limit lines. |
SEGMENTS ARE |
|
NOT ALLOWED |
|
NO EXTERNAL | Attempting to use the Bias Tee option without external power applied |
POWER TO BAT- | to the battery recharge port. |
TERY |
|
RECHARGER |
|
PORT. BIAS TEE |
|
CANNOT BE ACTI- |
|
VATED. |
|
BIAS TEE OPTION | Attempting to use the Bias Tee option when the option has not been |
NOT INSTALLED | installed. |
|
|