Setting the Amplitude Scale

Chapter 4 Measurements

Step 15. Remove the Short, connect the calibrated Load and press the ENTER key. The message “MEASURING LOAD ” will appear.

NOTE:

For OSLT calibration, the messages “MEASURING LOAD” and “CONNECT LOADS TO RF OUT AND RF IN PORTS” will appear.

Step 16. Connect the second calibrated load to the RF In test port and press the ENTER key. The messages “MEASURING ISOLATION” and “CONNECT RF OUT PORT TO RF IN PORT” will appear.

Step 17. Remove both loads, connect an adapter and/or a cable from the RF Out test port to the RF In test port and press ENTER.

Step 18. Verify that the calibration has been properly performed by checking that the message “CAL ON” is displayed in the upper left corner of the display.

Setting the Amplitude Scale

The following procedure sets the amplitude scale.

Step 1. Press the AMPLITUDE or LIMIT key to call up the Scale menu.

Step 2. In SWR mode, press the TOP soft key and enter 1, ., 1 using the key pad or the Up/Down arrow key, and press ENTER. (In Return Loss mode, enter 0; in Inser- tion Gain/Loss mode, enter 100.)

Step 3. In SWR mode, press the BOTTOM soft key and enter 1 using the key pad or the Up/Down arrow key, and press ENTER. (In Return Loss mode, enter 54; in In- sertion Gain/Loss mode, enter –120.)

NOTE:

Figure 4-3and 4-4show the typical responses of a valid OSL calibration with a load attached to the test port with 517 data points.

Figures 4-5,4-6,4-7and 4-8show typical responses of a valid OSLT calibra- tion with loads attached to the test port and with the RF Out test port attached to the RF In test port.

To change the number of data points, press the SWEEP key, followed by the RESOLUTION soft key. Select 130, 259, or 517 data points.

Saving the Setup

The following procedure can be used to save a setup and calibration in memory.

Step 1. Press the SAVE SETUP key.

Step 2. Use the key pad or the Up/Down arrow key to select a memory location (1 - 10), then press ENTER.

Recalling a Setup

The following procedure recalls a setup from memory.

Step 1. Press the RECALL SETUP key.

4-4

Page 50
Image 50
Anritsu S251B manual Setting the Amplitude Scale, Saving the Setup, Recalling a Setup, Measurements