Chapter 4 Measurements
Step 7. Use the Up/Down arrow key to scroll to F1, then press ENTER to select the start frequency F1.
Step 8. Enter 6, 2, 5 using the keypad or the Up/Down arrow key. Press ENTER to set F1 to 625 MHz.
Step 9. Use the Up/Down arrow key to scroll to F2, then press ENTER to select the start frequency F2.
Step 10. Enter 2, 5, 0, 0 using the keypad or the Up/Down arrow key. Press ENTER to set F2 to 2500 MHz.
Step 11. Scroll to PROP VEL and press ENTER to select Propagation Velocity parameters.
NOTE:
Propagation velocity and cable loss can be accessed by selecting CABLE FUNCTION from the DTF AID parameter table.
Step 12. Enter ., 8, 6 and press ENTER to set the Propagation Velocity to 0.86.
Step 13. Press ENTER to start the calibration.
NOTE:
If the calibration is valid, scroll to CALIBRATION
Performing a Calibration
This procedure is used to perform a calibration.
Step 14. At the Perform Calibration menu, follow the instructions on the display to con- nect the respective calibration components to the Site Master test port. (Refer to the calibration procedure on page
NOTES:
For best calibration results — compensation for all measurement system uncer- tainties — ensure that the Open/Short/Load is at the end of the test port or op- tional extension cable; that is, at the same point that you will connect the antenna or device to be tested.
For best results, use a phase stable Test Port Extension Cable (see Optional Accessories). If you use a typical laboratory cable to extend the Anritsu test port to the
For optimum calibration, Anritsu recommends using precision calibration com- ponents.