Anritsu S251B manual 4-27/4-28, Making Power Measurements, Entering Power Monitor Mode

Models: S251B

1 84
Download 84 pages 19.43 Kb
Page 73
Image 73
Making Power Measurements

Chapter 4 Measurements

Making Power Measurements

Power measurement is accomplished using a broadband (1 MHz to 3000 MHz) RF detector, P/N 5400-71N50. The power monitor displays measured power in dBm or Watts.

Entering Power Monitor Mode

Step 1. Press the MODE key.

Step 2. Use the Up/Down arrow key to scroll to POWER MONITOR. Press ENTER to select power monitor mode.

Zeroing the Power Monitor

Step 3. With no power applied to the DUT, press the ZERO soft key from the Power menu. Wait a few seconds while the Site Master accumulates samples of the qui- escent power level. When complete, ZERO ADJ: ON is displayed in the message area.

Measuring High Input Power Levels

Step 4. Insert an attenuator between the DUT and the RF detector, sufficient to insure that the input power to the Site Master is no greater the 20 dBm.

Step 5. Press the OFFSET soft key.

Step 6. Enter the attenuation in dB using the keypad. Press ENTER to complete the en- try. The message area will show OFFSET is ON along with the entered value in dB.

Displaying Power in dBm and Watts

Step 7. Press the UNITS soft key to display power in Watts.

Displaying Relative Power

Step 8. With the desired base power level input to the Site Master, press the REL soft key. The message area will show REL: ON and the power reading will indicate 100%.

Step 9. Press the UNITS soft key to display power in dBm. Since REL is ON, the power reading will be in dBr, relative to the base power level.

4-27/4-28