Chapter 4 Measurements
Step 3. Press the ENTER key to select the measurement mode.
| LOAD |
| OPEN |
TEST PORT CABLE | SHORT |
(OPTIONAL) | LOAD |
|
TEST PORT CABLE (OPTIONAL)
RF IN TEST PORT
RF OUT TEST PORT
Site Master S251B
1 | 2 |
START AUTO
CAL SCALE
3 4
ESCAPE
CLEAR
|
|
| SAVE | RECALL |
|
|
|
| SETUP | SETUP |
|
|
|
| 5 | 6 |
|
|
|
| LIMIT | MARKER |
|
|
|
| 7 | 8 | ENTER |
|
|
|
| ||
|
|
| SAVE | RECALL | RUN |
|
|
| DISPLAY | DISPLAY | HOLD |
|
|
| 9 | 0 | +/- |
625.0 |
|
| 2500.0 |
|
|
1384.45 MHz |
|
|
|
| |
2096.66 MHz |
| ON | SYS | ||
|
|
| OFF | . | |
MODE | FREQ/DIST | AMPLITUDE | SWEEP |
|
|
Figure
Selecting the Frequency Range
The following procedure selects the frequency range for the calibration.
Step 4. Press the FREQ/DIST key.
Step 5. Press the F1 soft key.
Step 6. Enter 6, 2, 5 or a
Step 7. Press ENTER to set F1 to 625 MHz or the
Step 9. Enter 2, 5, 0, 0 or a
Step 10. Press ENTER to set F2 to 2500 MHz or the
Step 11. Check that the FREQ (MHz) scale in the display area indicates the new fre- quency start and stop values (for example: F1 = 625 MHz, F2 = 2500 MHz).
Performing a Calibration
The following procedures detail how to perform the calibrations. Steps 12 through 15 de- scribe OSL calibration, while steps 16 and 17 apply only to OSLT calibration.
Step 12. Press the START CAL key. The message “CONNECT OPEN TO RF OUT PORT” will appear in the display.
Step 13. Connect the calibrated Open and press the ENTER key. The messages “Mea-
suring OPEN” and “CONNECT SHORT TO RF OUT” will appear.
Step 14. Remove the Open, connect the calibrated Short and press the ENTER key. The messages “MEASURING SHORT” and “CONNECT LOAD TO RF OUT” will appear.