Operation | MAC2000 |
4.Depending on the percentage of air found in the internal tank, perform one of the following:
•If the display shows “Identifier result: EXCESSIVE,” then purge the tank by first pressing the Reset key to display the message “Would you like to purge air?” Then press the Start key to begin purging air. When complete, the Main menu will appear. Repeat Steps 2, 3, and 4 until the percentage of air within the internal tank is at an acceptable level.
•If the display shows “Identifier result: RECOVERABLE,” then the amount of air within the tank is at an acceptable level. Press the Reset key to display the message “Would you like to purge air?” Then press the Reset key again to return to the Main menu.
28 | Instruction |