Bacharach 2072-0400 manual 4.4Replacing the Refrigerant Identifier Filter

Models: 2072-0400

1 40
Download 40 pages 52.22 Kb
Page 36
Image 36
4.4Replacing the Refrigerant Identifier Filter

Maintenance

MAC2000

4.4Replacing the Refrigerant Identifier Filter

Caution: Visually inspect the Refrigerant Identifier Filter every day. If it begins to turn red, replace it immediately! You risk damaging the identifier sensor if the filter is not replaced.

The refrigerant identifier filter is located on the control panel, directly under the Low Side pressure gauge. The filter’s function is to protect the refrigerant identifier sensor from contaminates. When this filter starts to develop red spots, replace the filter immediately!

1.Remove the filter from its holder; then remove the tubing from each end.

2.Position the new filter so that its flow-arrow points in the same direction as the flow-arrow on the control panel.

3.Attach the tubing to the new filter, and press the filter back into its holder.