2.4Probe Zero & Calibration

CMX High Performance Material & Coating Thickness Gauge

3)Press the ENTER key to select the transducer type and display overwrite existing probe screen.

4)Press the OK key to overwrite the existing probe type with the newly selected probe type. The zero probe screen will be displayed. Proceed to the zero probe section that follows.

2.4Probe Zero & Calibration

The next steps are to perform a probe zero and calibrate the CMX to the material and transducer being used. If the sound velocity is unknown, the CMX can be calibrated to a known thickness sample. This demo will briefly explain both of these techniques.

The CMX is equipped with two zero options:

1)Off Block Zero (Automatic Probe Zero) – When this feature is enabled the CMX will do an electronic zero automatically, eliminating the need for a zero disk or block.

2)On Block Zero (Manual Probe Zero) – When this feature is enabled the transducer must be placed on the probe zero disk (battery cover located on the top of the unit.

Note: Transducers of the same type will have very slight mechanical and electrical variations. If it’s discovered that the linearity is off following an initial auto probe zero and extreme accuracy is required, a manual zero should be performed followed by an auto zero. This will adjust and eliminate any error. This is only required if it’s discovered the transducer is non-linear following an initial auto probe zero.

The procedures are outlined as follows: