Dakota Digital CMX operation manual One Point Calibration

Models: CMX

1 104
Download 104 pages 54.3 Kb
Page 81
Image 81
Manual background

CMX High Performance Material & Coating Thickness Gauge

One Point Calibration

Note: Use the maximum coating sample for the one point calibration first.

1)Physically measure the thicker of the two samples of coating, as close as possible to the maximum expected coating measurement range, using a set of calipers or a digital micrometer.

Very Important: If coating measurements will be made with the coating applied to a metal surface, the calibration must be done in the same manner, with the samples coupled to a metal surface. However, if the coating will be measured as a stand alone material, the calibration must be performed the same way.

2)Apply a drop of couplant on the transducer and place the transducer in steady contact with the thicker of the two coating and samples fist. Be sure that the reading is stable and the repeatability indicator, in the top left corner of the display, is fully lit and stable. Press the MENU key once to activate the menu items tab. Press the MENU key multiple times to tab right and the ESC key multiple times to tab left until the CAL menu is highlighted and displaying the submenu items.

77

Page 81
Image 81
Dakota Digital CMX operation manual One Point Calibration