In this first example the transducer was automatically identified:

Probe Automatically Recognized

1)Press the OK key once to use the identified probe, or ESC to display a list of optional transducers. Note: if the CMX recognizes a specific transducer, the user should always select OK to use the identified probe. The only time an alternative probe should be selected from a list is if the user switched probes following initial power up and recognition, or the CMX has somehow identified the probe in error..

2)Assuming the CMX recognized the probe and the OK key was pressed, the CMX will advance to a Zero Probe menu. If the transducer was identified as a special transducer capable of measuring coating thickness, a menu will be displayed allowing the user the ability to toggle the coating thickness display on/off as follows: