2.6 Coating Calibration

Dakota Ultrasonics

6)Use the UP and DOWN arrow keys to scroll through the sub menu items until ZERO COATING is highlighted.

7)Press the ENTER key to display the confirmation screen.

8)Press the OK key to zero the coating and retur n to the PRB menu, or ESC to cancel the coating zero process.

9)Press the MULTI MODE key once to activate the measurement mode options.

10)Use the UP and DOWN arrow keys to scroll through the sub menu items until Coating On (PECT) is highlighted.

11)Press the ENTER key to select the measurement mode and return to the measurement screen, and begin taking readings.

2.6 Coating Calibration

The CMX has been preset to a default coating velocity of 0.0850 in/µsec (2159 m/sec). This will be very close to the most common coating material velocities used in the field. If the velocity of the coating is known, and different than the above default setting, the user can simply enter the coating velocity into the CMX. However, if the velocity is unknown, the CMX can also be calibrated to a specific coating sample/type using the 1pt calibration option in PECT (pulse-echo coating) mode, or a two point calibration is CT (coating only) mode. For the purpose of this