MMX-6 DL Ultrasonic Thickness Gauge

The Back light

The back light of the MMX-6 DL can be set to three different positions on/off/auto. The following steps outline the procedure for changing the settings of the back light:

Using The Back Light

1)Press ON/OFF key to power up the MMX-6 DL.

2)Press the MODE key to activate features and settings.

3)Press the UP / Down arrow keys to scroll to the LItE symbol.

4)Press the SEND key to toggle the status of the back light on/off/auto.

5)Press the MODE key once again to return to measurement mode.

Scan Mode

While the MMX-6 DL excels at making single point measurements, it is sometimes desirable to examine a larger region, searching for the thinnest point. The MMX-6 DL includes a feature, called Scan Mode, which allows it to do just that.

In normal operation, the MMX-6 DL performs and displays four measurements every second, which is quite adequate for single measurements. In Scan Mode, however, the tool performs sixteen

23

Page 27