Dakota Digital operation manual MMX-6 DL Modes Changing Units in / MM, Units in/mm

Models: MMX-6

1 49
Download 49 pages 16.05 Kb
Page 26
Image 26

Dakota Ultrasonics

8)Press the Probe key. The display will flash 1OF2. Repeat steps 3 through 7 on the second calibration point. The MMX-6 DL will now display the sound velocity value it has calculated based on the thickness values that were entered in step 7. The MMX-6 DL is now ready to perform measurements within this range.

MMX-6 DL Modes

Changing Units IN / MM

The MMX-6 DL has the ability to display measurements in both the English (IN) and Metric (MM) systems. The following steps outline the procedure for changing units:

Units in/mm

1)Press ON/OFF key to power up the MMX-6 DL.

2)Press the MODE key to activate features and settings.

3)Press the UP or Down arrow keys to scroll to the unIt symbol.

4)Press the SEND key to toggle the status of the units - IN / MM.

5)Press the MODE key once again to return to measurement mode.

22

Page 26
Image 26
Dakota Digital operation manual MMX-6 DL Modes Changing Units in / MM, Units in/mm