Emerson 2400S installation manual 1 Rotating the transmitter on the sensor

Models: 2400S

1 50
Download 50 pages 13.6 Kb
Page 10
Image 10
Figure 2-1Rotating the transmitter on the sensor

Transmitter Orientation and Power Supply

Figure 2-1Rotating the transmitter on the sensor

Clamping ring

Feedthrough

Feedthrough notches

Manual backgroundManual backgroundManual background CAUTION

Excessive rotation of the housing on the feedthrough can cause wiring damage and result in measurement error or flowmeter failure.

To reduce the risk of damaging internal wiring, do not rotate the housing more than 360°. You can rotate the transmitter either clockwise or counter-clockwise to reach the desired position.

2.4Rotating the user interface module on the transmitter (optional)

For easier access, the user interface module (whether or not the transmitter has an LCD display) can be rotated on the transmitter up to 360° in 90° increments.

To rotate the user interface module on the transmitter:

1.Remove power from the unit.

Manual backgroundManual backgroundManual background WARNING

Removing the transmitter housing cover in a hazardous area while the transmitter is powered up can cause an explosion.

To avoid the risk of an explosion, remove power from the transmitter before removing the transmitter housing cover.

2.Referring to Figure 2-2,remove the transmitter housing cover and user interface module:

a.Loosen the four transmitter housing cover screws.

b.Remove the transmitter housing cover.

c.Loosen the two user interface screws.

d.Gently lift the user interface module, disengaging it from the user interface connector on the transmitter.

6

Micro Motion® Model 2400S Transmitters

Page 10
Image 10
Emerson 2400S installation manual 1 Rotating the transmitter on the sensor