Emerson 2400S Environmental limits, A.10 Environmental effects, A.11, All models, Specifications

Models: 2400S

1 50
Download 50 pages 13.6 Kb
Page 43
Image 43
Environmental limits

Dimensions and Specifications

A.9

Environmental limits

 

 

 

Ambient temperature limits

Operating and storage: –40 to +140 °F (–40 to +60 °C)

 

 

Below –4 °F (–20 °C), LCD responsiveness decreases and LCD may become

 

 

difficult to read. Above 131 °F (55 °C), some darkening of the LCD panel might

 

 

occur.

 

 

ATEX requires limiting ambient temperature to below 131 °F (55 °C).

 

 

Humidity limits

5 to 95% relative humidity, non-condensing at 140 °F (60 °C)

 

 

Vibration limits

Meets IEC68.2.6, endurance sweep, 5 to 2000 Hz, 50 sweep cycles at 1.0 g

 

 

 

A.10 Environmental effects

All models

 

EMI effects

Meets EMC directive 89/336/EEC per EN 61326 Industrial

 

Conforms to NAMUR NE21 Version: 10.02.2004

 

 

Model 2400S Analog only

 

Ambient temperature effect

On mA output: ±0.005% of span per °C

 

 

A.11

Hazardous area classifications

 

 

 

All models

 

CSA(1) C-US

Class I Division 2 Groups A, B, C, D

 

 

Class II Division 2 Groups F and G

ATEX(2)

 

 

Zone 1 Analog

II 2GD EEx mb e ib IIC T4

 

 

 

 

Zone 2 Analog or

II 3G EEx n A C II T5

 

PROFIBUS-DP

II 3D IP66/IP67 T70°C

 

 

 

 

 

 

Zone 2 DeviceNet

II 3G Ex nA II T5

 

 

II 3D IP66/IP67 T70°C

 

 

 

(1)CSA is a Canadian approvals agency that provides approvals accepted both in the U.S.A. and in Canada.

(2)ATEX is a European directive.

Specifications