Configuring the 268XA

Reviewing and Setting the General Network IP Address

Perform the procedure below to review or set the device’s general network Internet Protocol (IP) address.

General Network IP Address Enter the IP Address supplied by your network administrator for each BCN. The format is four 3 digit segments: IP0.IP1.IP2.IP3.

Table 6. Reviewing and Setting the General Network IP Address

￿Press the COMM key to review the network settings, or press and hold the COMM key for 3 seconds to set the network settings.

￿

￿

￿

￿

￿

￿

Press the up/down arrow keys until nEt (Network) appears in the primary

display (comm appears in the secondary display).

Press the ENTER key. nEt appears in the secondary display and ISo (isolated network) or gEn (general network) is in the primary display.

To set the network type, press the up/down arrow keys to

select gEn.

Press the ENTER key. (Pressing any other key will cancel set operations.) This displays the current Socket Port.

￿Press the ENTER key. This displays the first digit of the 12 digit Internet Protocol address (grouped into four 3 digit segments: IP0 . IP1 . IP2 . IP3).

￿

￿

￿

￿

￿

Press the left/right arrow keys to select the desired number in each segment.

The selected digit is highlighted and the segment, for example, IP2, appears.

Press the up/down arrow keys to select the desired number, 0 to 9, for the

positioned IP digit. In this manner, select all 12 IP digits.

Press the ENTER key to enter the settings and exit the procedure. (Pressing any other function key will cancel set operations.).

13

Page 23
Image 23
Fluke 2686A, 2680A manual Reviewing and Setting the General Network IP Address

2686A, 2680A specifications

The Fluke 2680A and 2686A are sophisticated data acquisition systems designed for versatile applications in electrical and environmental testing. These high-performance devices cater to engineers and technicians who require reliable and precise measurement capabilities.

At the heart of these instruments is their advanced modular design, which allows users to customize the system according to specific measurement needs. The Fluke 2680A offers eight channels, while the 2686A boasts an impressive 20 channels of data acquisition, thus providing flexibility for various testing scenarios. This modularity ensures that users can easily expand their measurement capabilities by adding more channels or switching between different measurement modules.

One of the standout features of the Fluke 2680A and 2686A is their ability to perform simultaneous measurements across multiple channels. This is particularly beneficial for applications that require real-time data analysis, minimizing the time spent on sequential measurements. With sample rates reaching up to 100 kS/s, these devices provide quick and accurate data collection, making them ideal for both laboratory settings and field use.

Both models incorporate advanced signal conditioning technology, enabling them to handle a wide range of input types, including voltage, current, temperature, and resistance. This capability eliminates the need for external signal conditioning equipment, streamlining the setup and reducing overall costs for users. The inclusion of built-in digital filtering and scaling further enhances measurement accuracy, ensuring reliable results even in noisy environments.

The Fluke 2680A and 2686A also feature robust connectivity options, including USB, Ethernet, and RS-232 interfaces. This promotes seamless integration into existing networks and facilitates remote monitoring and control. Additionally, both devices support various programming languages, allowing for easy automation of testing procedures.

Another highlight is the intuitive user interface that allows users to configure the device with minimal effort. An embedded web server provides remote access to the instrument, enabling users to monitor data and configure settings from anywhere.

In summary, the Fluke 2680A and 2686A stand out as premier data acquisition systems, offering exceptional modularity, versatile measurement capabilities, and advanced connectivity options. Their combination of speed, accuracy, and ease of use makes them indispensable tools for professionals in testing and measurement applications.